- 22 -
DeviceNet
Messaging
Poll Command Message:
Byte
7
6
5
4
3
2
1
0
0
PUMP
PUMP = Command for pump to be on or off. 1 = on, 0 = off..
Respond Message:
Byte
7
6
5
4
3
2
1
0
0
Water Temperature (LSB)
1
Water Temperature (MSB)
2
Water Resistivity (LSB)
3
Water Resistivity (MSB)
4
Flow 1 (LSB)
5
Flow 1 (MSB)
6
Flow 2 (LSB)
7
Flow 2 (MSB)
8
Flow 3 (LSB)
9
Flow 3 (MSB)
10
Flow 4 (LSB)
11
Flow 4 (MSB)
12
Flow 5 (LSB)
13
Flow 5 (MSB)
14
Flow 6 (LSB)
15
Flow 6 (MSB)
16
Flow 7 (LSB)
17
Flow 7 (MSB)
18
Flow 8 (LSB)
19
Flow 8 (MSB)
20
Facility Water Flow (LSB)
21
Facility Water Flow (MSB)
22
AF WLW PUMP FW CB WLF
23
0
0
0
0
0
0
0
0
Upon receipt of a poll command, we will send the 23 bytes of data through the
DeviceNet. The 23 byte response is in accordance to the Device Net fragmen-
tation protocol of the DeviceNet Specification (revision 1.3).
CB = Cartridge Good/Bad, 1 = Good, 0 = Bad
WLF = Water Level Low Fault, 1 = Good, 0 = Bad
FW = Facilities Water Flow, 1 = Good, 0 = Bad
PUMP=Status of Pump, 1 = ON, 0 = OFF
AF = Autorefill Status, 1 = ON. 0m = OFF
WLW = Water Level Warning, 1 = OK, 0 = OFF