Thermo Scientific Robot Stainer Microm HMS 740

 

Microm International GmbH 

part of Thermo Fisher Scientific 

Otto-Hahn-Str. 1A 

69190 Walldorf / Germany 

 

387801 - English 

1

Thermo Scientific  
Robot Stainer  
Microm HMS 740 
387801 Issue 10 
 
INSTRUCTION MANUAL 

Содержание Microm HMS 740

Страница 1: ...t Stainer Microm HMS 740 Microm International GmbH part of Thermo Fisher Scientific Otto Hahn Str 1A 69190 Walldorf Germany 387801 English 1 Thermo Scientific Robot Stainer Microm HMS 740 387801 Issue...

Страница 2: ...Thermo Scientific Robot Stainer Microm HMS 740 Microm International GmbH part of Thermo Fisher Scientific Otto Hahn Str 1A 69190 Walldorf Germany 387801 English 2...

Страница 3: ...d or replaced free of charge by the manufacturer No other warranty is expressed or implied Unauthorized modification or repair by third party persons will void the warranty The warranty will expire in...

Страница 4: ...r All warranties and guarantees would then be null and void Only skilled or specially trained personnel must operate the slide stainer i e staining of histological and cytological specimens The listed...

Страница 5: ...ry default settings 2 3 Technical data Robot Stainer HMS 740 PART 3 OPERATING INSTRUCTIONS 3 1 Setting up the slide stainer 3 2 Initial turn on 3 2 1 Turning on and or off the instrument 3 2 2 Log in...

Страница 6: ...RT 4 THEORY OF OPERATION OF THE ROBOT STAINER HMS 740 4 1 Structural design 4 2 Operating features 4 3 Function PART 5 WORKING WITH THE ROBOT STAINER HMS 740 5 1 Preparations 5 2 Staining of specimens...

Страница 7: ...er Special warning messages to prevent harm to persons and or serious damage to equipment For your own safety please observe these instructions carefully Biohazard Warning of biological danger Radioac...

Страница 8: ...d must meet the International Electrotechnical Commission IEC regulations CAUTION MAINS VOLTAGE Never remove instrument covers during operation Component replacements as well as adjustments must only...

Страница 9: ...ctious material appropriate safety and disinfection measures must be carried out According to the rules and regulations concerning the handling of infectious material of the respective laboratory safe...

Страница 10: ...obot Stainer HM 420 must solely by used for staining histological cytological and hematological stainings Only skilled or specially trained personnel must carry out staining of histological cytologica...

Страница 11: ...ot Stainer HMS 740 is equipped with three transportation locks for the respective moving axes X Y and Z axis Danger Before the initial turn on of the instrument it is absolutely necessary to remove th...

Страница 12: ...A 69190 Walldorf Germany 387801 English 12 Transportsicherungen vor Inbetriebnahme entfernen Remove the transportation locks before the initial turn on Enlever les verrouillages de transport avant la...

Страница 13: ...ing and the swing open acrylic glass cover allow and easy access to the bath containers The bath containers can be removed individually or in groups of four and filled as needed The loading station an...

Страница 14: ...signals in case of malfunctions Several bath layouts and up to 50 protocols with up 100 protocol steps each can be stored Immersion time can individually be selected between 1 sec up to 99 min 59 sec...

Страница 15: ...tional GmbH part of Thermo Fisher Scientific Otto Hahn Str 1A 69190 Walldorf Germany 387801 English 15 TOTAL OVERVIEW ROBOT STAINER HMS 740 Fig 1 1 Hood 2 Touch screen 3 Quick start buttons 4 Bath con...

Страница 16: ...ower cord 220 240 V 50 60 Hz 100 120 V 50 60 Hz 1 water filter 1 feed hose 3 m inner diameter 13 mm thickness 3 5 1 drain hose 3 m inner diameter 19 mm thickness 4 1 charcoal filter already installed...

Страница 17: ...17 2 2 2 ADDITIONAL ACCESSORIES OPTIONAL Cat No Reagent container 556240 Cover 425780 Charcoal filter 184000 Standard basket for CoverTech 30 172550 Basket carrier for standard baskets 556200 Basket c...

Страница 18: ...of Thermo Fisher Scientific Otto Hahn Str 1A 69190 Walldorf Germany 387801 English 18 2 2 3 FACTORY DEFAULT SETTINGS Feature Setting Standard passwords User routine user admin Supervisor admin Note Wh...

Страница 19: ...conds Transportation storage conditions Temperature range 20 C up to 50 C Water pressure between 2 5 and 4 bar DI water pressure between 0 2 and 4 bar Operating conditions 5 C up to 35 C at a max rel...

Страница 20: ...safe transportation please fasten the four handles on the instrument Carry the instrument to the site of installation and place it on an even surface Danger Before the initial turn on first remove th...

Страница 21: ...nnections between water supply and instrument For safety reasons an aqua stop device as it is used for household instruments should be installed in addition Enclosed drain hose inner 19 mm length 2 5...

Страница 22: ...t not be so high that the specimens are removed from the slides Optional interfaces An external keyboard can optionally be connected to the USB interface fig 3a 1 An external mouse can optionally be c...

Страница 23: ...nts for the operation of the instrument Warning Before turning on the instrument for the first time please check if the power requirements indicated on the type plate fig 5 correspond to the power sup...

Страница 24: ...ode The instrument can now be started via the green button fig 7 1 on the touchpad keyboard Hinweis After the boot process of the computer approx 20 sec is finished an initialization phase of the indi...

Страница 25: ...the touch screen The user name and the password can be entered here For the initial turn on Name fig A and the the word admin must be entered and confirmed with OK Enter the password admin into the d...

Страница 26: ...lowed to carry out all settings Users can be deleted from a table For this tip the line with the concerned user on the left side of the table with your finger Confirm the process in the appearing dial...

Страница 27: ...ough the program sequences Main screen In the lower third of the screen the bath layout as well as the function bar are shown The bath layout fig 8 1 is shown as follows Looking from the front the fir...

Страница 28: ...to run down the computer in a controlled way due to a PC or software problem the emergency shutdown can be activated Here it is possible to run down the computer without destroying or losing necessar...

Страница 29: ...dy entered and cannot be altered Bath type The kind of station appears automatically Exception The stations 16 27 and 33 36 can be defined as Waste stations Name Enter the bath name here by selecting...

Страница 30: ...der USED click RESET Time uncritical To guarantee an optimal and efficient working with the instrument the first step in a staining protocol can be an uncritical one See 3 4 3 Reagent list To simplify...

Страница 31: ...re of the heated reagent stations can be selected between 30 C and max 75 C in 1 C increments Danger The temperature of the surfaces of the reagent stations can be increased For this please observe th...

Страница 32: ...69190 Walldorf Germany 387801 English 32 3 4 2 DELETING A BATH LAYOUT In case a bath layout is no longer used it can be deleted Press the button menu Select the submenu BATHS Then select DELETE BATH L...

Страница 33: ...e stainer is thus ensured After having started a protocol the baskets are immediately transported from the entry station into the time uncritical baths Caution Time uncritical baths can only be define...

Страница 34: ...not emptied into the rinsing station however it is emptied into the programmed bath container This bath container can be removed once a day to dispose of the contents appropriately and according to th...

Страница 35: ...n be deleted within the bath layout For this open the menu Baths under Edit bath layout Now select the waste station within the bath layout Select Bath under Bath type This station can now be provided...

Страница 36: ...LOAD With NEW a completely new protocol can be processed The following entries must be made for a new protocol in the field Stepdata Step The corresponding step is selected by choosing the respective...

Страница 37: ...or message is shown After having created a protocol select SAVE AS Enter a protocol name and confirm with Enter Note A certain bath layout can be assigned to a protocol Select SELECT in the field BATH...

Страница 38: ...l and confirm with OK Via the field Stepdata alterations can be made For this select the step which is to be altered via the arrow buttons and enter the alteration In case a step is to be added this s...

Страница 39: ...Otto Hahn Str 1A 69190 Walldorf Germany 387801 English 39 3 5 3 DELETING STAINING PROTOCOLS To delete staining protocols which are no longer used please proceed as follows Select the menu button Then...

Страница 40: ...lose the hood Open the entry station and remove the cover from the station Hook in a basket carrier with a basket Close the drawer The Start button on the touch screen is now green In case a protocol...

Страница 41: ...warning signal This acoustic message can be acknowledged by activating the O K button on the touch screen Moreover the error message EXIT OCCUPIED is shown on the screen Caution Remove one of the thre...

Страница 42: ...prox 5 sec Within this time a protocol which is to be cancelled can be selected For this directly tip onto the protocol on the touch screen A cancellation window opens up on the screen gray background...

Страница 43: ...WITH EQUAL NAMES WILL BE OVERWRITTEN ON FLOPPY DISC The message can be confirmed with OK or cancelled via CANCEL When all dates are stored to the disc the following message appears on the screen BACKU...

Страница 44: ...OPPY DISC This message can be confirmed with OK or cancelled with CANCEL When all dates are stored on the disc the following message appears on the screen BACKUP TO FLOPPY DISC COMPLETED Restoring bat...

Страница 45: ...ransportation times exact immersion times in the baths bath status how often was a bath used error messages Note A QM report is deleted after 30 days without further queries in advance QM reports can...

Страница 46: ...ton Then select the button QM Select PRINT QM REPORTS and select the reports which are to be printed Now click on the button PRINT The selected reports are then stored onto the disk as pdf files Note...

Страница 47: ...itation Clock Sound For this select the menu button on the touch screen Then select SETTINGS Select USER SETTINGS The window with the user settings opens Language selection The desired language can be...

Страница 48: ...ed Setting the dripping movement To minimize the reagent carryover the number of shake movements can be selected from 0 5 After lifting a basket carrier out of the reagent bath the adherent reagent is...

Страница 49: ...n be selected here After having carried out all desired settings confirm them with OK The following message will appear CHANGES MADE IN THIS DIALOG WILL TAKE EFFECT ONLY AFTER THE RESTART OF THE MACHI...

Страница 50: ...English 50 3 10 2 INSTALL UPDATE Note Please observe the installation instructions on the supplied USB stick Before installing an update the instrument must be shut down and turned off and on again o...

Страница 51: ...e controlled and filled with with fresh water The water is drained into a water basin which has the size of approx the entire base of the instrument for safety reasons The waste waters as well as the...

Страница 52: ...emptied via a drain faucet into a rinsing or waste station at regular intervals This avoids the spilling of liquids The emptied liquids are then washed out of the respective rinsing station When using...

Страница 53: ...strument please note the following Check that all stations including exit stations are covered with a cover Check if the liquid level in each station is correct max filling height for staining solutio...

Страница 54: ...ion Close the drawer Select a protocol either via Start Protocol on the touch screen or via a quick start button Note If the protocol is to be started in the first step press the quick start button tw...

Страница 55: ...The drawer for the entry and exit stations must be closed By no means open it while the basket carrier is just inserted or removed from a station Check if the charcoal filter has been inserted Danger...

Страница 56: ...protective switch was activated due to an error Contact your Thermo Scientific supplier No water flow No water connection Check water connection Check pressure controller Contact Thermo Scientific sup...

Страница 57: ...ries of persons keep and close the hood on its handle Remove the bath covers and clean them with appropriate agents outside the instrument Remove the bath containers and dispose of the contents accord...

Страница 58: ...ffers a service contract which guarantees that your instrument is always in perfect condition For more information please contact the nearest Thermo Scientific sales office Note We strongly recommend...

Страница 59: ...erating hours a message is shown on the screen and asks for the replacement of the charcoal filter Loosen the screws of the cover on the right side of the bath stations Remove the cover Loosen and rem...

Страница 60: ...rument Unplug the unit Remove covers and other accessories from the instrument Loosen the water connections Insert the transportation safety device for the individual movement axes Install the carryin...

Страница 61: ...al shut down of the instrument we recommend to contact a local recycling company for the disposal according to the national applicable regulations To be applied in the countries of the European Union...