10
Routine Cleaning
Clean up any spills immediately. Remove the plug from its
outlet for a thorough cleaning between incubation periods.
The chamber interior and shelves can be cleaned with a
mild detergent or soap and water. Do not use abrasives or
metallic scouring pads on any part of the unit.
To clean the viewing door use a soft damp cloth. A regular
window cleaning solution can be used. Avoid scratching
the acrylic plastic.
Repairs and Servicing
Unplug the unit from its outlet before servicing. The ther-
mostat is easily accessible inside the chamber.
A heater is located under the floor of the chamber. Turn
the incubator upside down and remove 4 screws from the
bottom of the cabinet. Lift out the bottom and the insula-
tion covering the heater. Disconnect the heater leads and
remove 2 nuts that hold the heater to the floor of the
chamber.
Maintenance
Warning
Disconnect plug from electrical outlet
before attempting any maintenance or
repair of this unit.
Note
Make no attempt to service or repair a
Thermo Scientific product under war-
ranty before consulting your distribu-
tor. After the warranty period, such
consultation is still advised, especially
when the repair may be technically
sophisticated or difficult.
If assistance is needed beyond what
the distributor can provide, please call
the customer service department at 1-
800-553-0039. No merchandise, how-
ever should be returned without prior
approval.