iCAP Customer Familiarisation Manual
Page 11 of 72
This document is part numbered 849940090071
Version 1.02
5.4 Centre tube insertion into torch holder
Insert the centre tube assembly into the metal torch holder.
Screw the centre tube holder assembly in a clockwise direction into the metal torch holder until
the o-ring is compressed. Do not over tighten as this will reduce the lifetime of the O-ring seal.
When fitted the center tube should be 1-3mm lower than the intermediate tube as shown above.
5.5 Torch holder insertion into the torch box
Insert the torch holder into the
torch box and turn the metal torch holder clockwise until the red orientation lock self locates in the
torch box casting.
5.6 Positioning of the spray chamber drain
Insert the white plastic tubing connector and wide bore tubing (0.79mm inner diameter) into the
spray chamber drain tube. The drain and spray chamber should be positioned so that no pulsing