iCAP Customer Familiarisation Manual
Page 13 of 72
This document is part numbered 849940090071
Version 1.02
5.8 Connection of spray chamber adaptor to torch assembly
Insert spray chamber adaptor fitting into torch assembly holder as far as it will go, and connect up
the Nebulizer gas supply to the push-fit fitting.
After assembly of the sample introduction system and prior to ignition of the plasma checks
should be made for correct assembly:
•
Make sure the torch is fully rotated and locked in place.
•
Make sure the centre tube holder is fully rotated and locked into the torch.
•
Make sure the spray chamber adaptor is fully pushed into the torch body.
•
Make sure the spray chamber is tightly clamped to the spray chamber adaptor.
Problems in any of these areas may cause air leaks or disruption of the gas flows making the
plasma difficult to ignite and may cause damage to the torch.
WARNING: It is extremely important that the correct Thermo Scientific parts are used for the
sample introduction system. In addition interlocks on the torch holder and other parts of the
instrumentation are there for safety and must not be bypassed. Operators could be exposed to
dangerous UV and radio frequency radiation if alternate parts are used for the spray chamber
adaptor.
▲