An iCAP 6000 Duo instrument will require an
extraction of 10 m/s
(33 ft/s; 22.4 mph)
.
An iCAP 6000 Radial instrument will require an
extraction of 5 m/s
(16 ft/s; 11.2 mph)
.
The extraction flow must be adjustable to 50% of the maximum to
allow optimization during installation. This can be achieved with a
butterfly valve, or an adjustable gate as shown in figure 4-3 below.
The extraction system must not be affected by external weather
conditions, or other uses the system may be used for.
Figure 4–4.
iCAP 6000 Extraction System
PLEASE NOTE: EXTRACTION
REQUIREMENTS ARE NOT THE SAME AS
FOR PREVIOUS THERMO FISHER ICP
SPECTROMETERS.
Measure flow rate at
instrument end of
extraction system