1.4 Pressure regulation
For safety, interlocks are recommended for pressure regulation inside
the BPC for all single-use systems. The gassing strategy should be
reduced as the pressure inside the BPC approaches 0.5 psi (0.03 bar).
Any time the pressure exceeds 0.5 psi (0.03 bar) in the BPC, the gas
flow should be stopped.
1.5 Motor cap safety sensor
An optional safety switch may be installed on the motor cap of all
single-use systems. This device uses a reed switch that is normally
closed when the cap is closed. When the cap is open, a safety
interlock should prevent the rotation of the drive shaft.