Thermo Scientific
Single-Use Systems Integrator’s Guide
|
13
Chapter 1
|
Configuration Guidelines
Please Note:
The impeller power number assumption of Np=2.1
is only an approximation, based on an academic exercise.
Recommended best practice for determination and comparison of
Np between bioreactors should be carried out by the end user and
should be based on conditions matching the specific application. End
users should also anticipate and accept some variability that may be
inherent in the analysis technique or modeling method chosen. If a
different or alternative Np value is determined by the end user, the
safety interlocks must be based solely on the maximum prescribed
RPM for a working volume as found in the reference tables in this
publication.
Never rely strictly on an estimated magnitude of
P/V when setting the agitator speed interlock levels.
Table 1.3. Maximum agitation rates (in RPM) for S.U.B.s at a P/V
of 40 W/m
3
. Using the 2,000 L S.U.B. above 20Wm
3
requires con-
sultation with Thermo Fisher Engineering.
50 L
S.U.B.
100 L
S.U.B.
250 L
S.U.B.
500 L
S.U.B.
1,000 L
S.U.B.
2,000 L
S.U.B.
Fill %
2.1 and 5.1 Systems
100
230
184
148
127
109
*
95
227
181
145
125
105
*
90
222
178
143
123
103
*
85
218
174
140
121
101
*
80
214
171
137
118
99
*
75
209
167
134
116
97
*
70
205
164
131
113
95
*
65
200
160
128
111
92
*
60
194
155
125
108
90
*
55
189
151
121
105
87
*
50
183
146
117
101
86
*
5.1 Systems Only
45
177
141
113
98
82
*
40
170
136
109
94
79
*
35
162
130
104
90
75
*
30
154
123
99
85
71
*
25
145
116
93
80
67
*
20**
135
107
86
75
64
*
* Consultation with Thermo Fisher Scientific engineers is required when operating 2,000 L
S.U.B.s at a P/V of 40 W/m
3
.
** RPM values assume the system is operating at no less than 20% volume. In order to
ensure proper volume measurements, end users are responsible for ensuring proper load
cell calibration prior to system use.
NOTE: Load cell values must be used with interlocks to keep the system operating
within recommended P/V ranges.