8.1 Units without Thermo Scientific
e-boxes
8.1.1 46 series heaters
The 46 series heaters should be plugged into an appropriate voltage
plug per the voltage rating of the heater. Adapters can be purchased to
convert the IEC C14 plug into the appropriate country-specific plug.
The 46 series heaters have an inline temperature control module, which
regulates the heater temperature to 55 °C. It is a factory-sealed non-
adjustable module.
Table 8.1. Pinout for 46 series vent filter heater plug
Item
Description
Thermo Scientific connector (male inline)
IEC C14
Integrator connector (female receptacle)
IEC C13
Connector manufacturer
Mouser
Recommended AWG
16/3 – SO-style cord
Connector Wiring
120 VAC
240 VAC
Pin E
Ground
Ground
Pin N
Neutral
Line 2
Pin L
Load
Line 1
Figure 8.3
. Pinout for 46 series
vent filter heater.
Thermo Scientific
Single-Use Systems Integrator’s Guide
|
50
Chapter 8
|
Venter Filter Heater Connection