background image

Monochromated X-Ray Source with XR5 Electron Gun 

Issue 5 

 

Monochromator and XR5 (HA600054)

11

To focus the X-rays from the source to a spot on the sample, the two quartz crystals 
must be formed to the correct geometry. This not only provides point to point focusing 
by mirror reflection but also maintains a constant angle of diffraction at the surface, 
which corresponds to the Bragg angle. 

Figure 3-4: A Schematic of Bragg X-Ray Diffraction 

The Bragg angle is critical to the monochromator performance and is related to the 
crystal lattice spacing and the wavelength of incident X-rays. Only X-rays of a certain 
wavelength incident to the crystal lattice structure at a specific angle will be reflected to 
the correct point; this angle is the Bragg angle. 

The ESCALAB 250 uses an apertured toroidal geometry on a 500 mm Rowland Circle 
to give the highest possible energy resolution. Each crystal substrate is machined to 
the exact form required and a thin single-crystal quartz wafer is then bonded to its 
surface.

This crystal assembly is then mounted in a precision mechanism as shown in Figure 
3.3. This assembly allows the orientation of the crystal to be set precisely, using two 
sets of three external micrometers without venting the system. 

The lower leg assembly provides an additional adjustment. This adjustment enables 
the “Bragg” angle between the anode and both crystals to be optimised; it is frequently 
referred to as the 'Leg Angle' as shown in Figure 3.5. 

Figure 3.5: Leg Angle Adjustment 

Monochromated X-Ray Source with XR5 Electron Gun

Monochromator and XR5 (HA600054) 

 

Issue 5  

12

3.5 

Moving the XR5 Anode 

If degradation of the aluminium anode does occur a new area of anode can be 
selected by moving the anode with the linear transfer drive mechanism until a 
satisfactory signal level is restored. The XR5 monochromator has been specifically 
designed so that the anode can be safely moved while the monochromator source is 
running.

The performance of the anode is most effectively checked by measuring the sensitivity 
for the Ag3d

5/2

XPS line on a sputter-cleaned standard silver sample in standard 

analyser modes. However when adjusting the anode position observation of either the 
spot on a phosphor sample or any high level XPS signal should suffice as a monitor. 

The linear motion drive moves the anode by 0.8 mm per turn. The range of motion and 
the useful anode length are both 25 mm. A fraction of a turn (for example a half turn) 
will usually be enough to expose a new aluminium film area. 

3.6 

Monochromator Fine Tuning 

Complete alignment and set up of standard operating conditions will generally be 
carried out by Thermo Fisher Scientific service engineers and most operators will not 
need to be concerned with any significant monochromator alignment. However 
monochromator fine-tuning is a procedure that both engineers and operators need to 
be familiar with. 

In order to maintain optimum performance of the ESCALAB 250 the instrument 
operator should check that the anode surface is in good condition on a daily basis, and 
renew this surface as described in the previous section. The operator should also 
regularly check (a minimum of once a week) that the analysis position (as defined by 
the microscope cross hairs and/or XP imaging) coincides with the true position of the 
X-ray beam on the sample. 

For these simple alignment checks it is essential to have a phosphor sample (such as 
Thermo Scientific Part No. SK1066-004) mounted on a sample holder, together with a 
piece of silver, or silver on silicon, and a TEM grid sample mounted over a hole. It is 
usually convenient to leave such a sample set-up mounted permanently on a sample 
block, and to leave it in the bottom of the FEAL when not required so that it remains 
free from contamination. 

3.6.1 

Alignment of the Analysis Centre and Microscope Cross Hairs 

Minor misalignments (<200 

P

m) of the microscope cross hairs and monochromator 

can develop over time. These can quickly be corrected without significantly affecting 
the instrument performance. 

Before making any adjustments to the monochromator ensure that the monochromator 
has had sufficient time to ‘warm up’, typically at least one hour. Frequently apparent 
minor misalignments of the monochromator spots are due to insufficient time being 
allowed for the gun to stabilise at the operating conditions. 

1.  Place a standard phosphor sample and a standard silver-on-silicon sample 

side by side on a sample holder. For convenience fit spacers under the thinner 
sample to make the sample heights equal. Ensure that the silver sample has a 
good earth contact. 

2.  Introduce these samples to the system. 

3.  Move the phosphor sample to the analysis position and check that the 

phosphor surface is in sharp focus on the microscope display monitor. 

4.  Start the mono source and set up a visible beam with the smallest spot size. 

Содержание ESCALAB 250

Страница 1: ...Moving the XR5 Anode 12 3 6 Monochromator Fine Tuning 12 3 6 1 Alignment of the Analysis Centre and Microscope Cross Hairs 12 3 7 T352 NT Computer Controlled Power Supply 14 4 Monochromator Operation 15 4 1 Avantage Data System Monochromator Controls 15 4 1 1 Errors and Warnings 16 4 2 Different XR5 Operating Modes 16 4 3 Switching on the Re circulated Water 17 4 4 Observing the Monochromator Spot...

Страница 2: ...ing the online manual loaded with the software datasystem It does not include detail on the initial installation and setting up of the electron gun or crystals nor any troubleshooting or maintenance which would require the user to directly access high voltage It includes procedures for replacement of certain consumable items such as anodes and filaments which require the system to be vented and su...

Страница 3: ...nterlocks are displayed in the software within the Trips Interlocks display as in GREEN OK or RED FAIL for the state of the interlocks Typical Trips Interlocks displays are shown in Figure 2 1 Occasionally pop up warning messages will appear After taking note of the message select the OK button Go into the details window and look at the Trips Interlocks you will see a red FAIL in the appropriate w...

Страница 4: ... software they include x Temperature in T352 Power Supply rises to unacceptable levels x Current drawn from T352 Power Supply HV greater than 60 mA x Drain from HV through X ray water lines greater than 20 mA Monochromated X Ray Source with XR5 Electron Gun Monochromator and XR5 HA600054 Issue 5 8 3 Description 3 1 Introduction The Monochromated X Ray Source produces X Rays which have well defined...

Страница 5: ...uit uses mains water or a re circulator as its input and then increases the flow through the X Ray source by use of a booster pump The control of the booster pump and of the solenoid valves which switch the water flow between the twin anode if fitted and mono source is by the datasystem via the T352 NT power supply Both the twin anode and mono anode are floating at high voltage and have water flow...

Страница 6: ...e by 0 8 mm per turn The range of motion and the useful anode length are both 25 mm A fraction of a turn for example a half turn will usually be enough to expose a new aluminium film area 3 6 Monochromator Fine Tuning Complete alignment and set up of standard operating conditions will generally be carried out by Thermo Fisher Scientific service engineers and most operators will not need to be conc...

Страница 7: ...ing Ratings Main HT 15kV 60mA Normal operating or 20kV 20mA Source conditioning Monochromator Filament 2 5A Bombard 3000V Cathode 1200V Focus 3500V Table 2 Specification Outputs for the Monochromator Gun Control is entirely from the computer no manual controls are fitted except for the rear panel mains power switch The supply requires five interlock loops Vacuum Water flow Twin anode cover Mono co...

Страница 8: ...the Apply button should then be selected Monochromated X Ray Source with XR5 Electron Gun Monochromator and XR5 HA600054 Issue 5 16 4 1 1 Errors and Warnings As the monochromator is ramping up to the desired operating point the data system will check for any problems and report these as a warning window If an Error does occur the data system will display a warning dialogue and instruct the operato...

Страница 9: ...upplied from the system or from a separate mains circuit in the laboratory If this is not switched on then when an attempt is made to switch on the monochromator the booster pump will run dry be much louder than normal and possibly be damaged The monochromator gun will not turn on as the water flow interlock will have failed 4 4 Observing the Monochromator Spot For fine tuning of the monochromator...

Страница 10: ...nd converge as the sample is raised and lowered 4 Position each spot by moving the dispersive and non dispersive adjusters on the crystal manipulator do NOT adjust the heights of the crystal as this will change the energy of the X Ray spot from each crystal to coincide with the cross hairs on the viewing camera These adjusters are very sensitive so only small adjustments should be made Never adjus...

Страница 11: ...eplacing Anodes In general the anode will need replacing only when all the aluminium coating has been degraded sufficiently to attenuate the X Ray flux below acceptable levels along its entire length If the anode condition requires its replacement the vacuum system will have to be fully shut down and vented to dry nitrogen Figure 5 3 XR5 Anode Assembly Refer to Figure 5 3 XR5 Anode Assembly 1 Unpl...

Страница 12: ... vented to dry nitrogen As a further precaution disconnect the mains supply from the T352 NT X Ray power supply Refer to the Figure 5 5 XR5 Lens and Emitter Assemblies 1 Remove the M8 x 16 hex head bolt clamping the XR5 earth strap 2 Unplug the leads from the XR5 3 Do not unbolt the 70mm FC38 11 pin feedthrough flange This is unnecessary except when replacing the feedthrough 4 Remove all the M8 x ...

Страница 13: ...ges Use a new FC64 copper gasket and make sure that the flange is offered to the gun in the correct orientation The main flange is labelled TOP on its circumference adjacent to the focus connection This should be in the top vertical position when refitted to the gun on the system 22 Check the pin base for filament and interlock continuity and all other pins for electrical isolation 23 Fully tighte...

Страница 14: ...Issue 5 28 5 9 Fault Finding Symptom General Cause Specific Cause How to Rectify How to Fix Mains re circulated water supply switched off Booster pump will make excessive noise Check supply Look for warning messages on datasystem Switch on supply Insufficient water flow Warning message on datasystem Water flow not sufficient to satisfy flow requirement Check inlet and outlet flow Look for warning ...

Страница 15: ...n still runs correctly Loss of control of the gun from the datasystem Gun still running An HT tick has effected communication Reboot the datasystem Try the gun control again Table 4 Fault Finding 5 10 Fuses Information for Service Engineers Within the digital X ray PSU T352 NT there is an 800mA T fuse on the mains input Monochromated X Ray Source with XR5 Electron Gun Monochromator and XR5 HA60005...

Отзывы: