Standardization & Calibration
Perform Calibration
Thermo Fisher Scientific
DensityPRO with
F
OUNDATION
™ fieldbus
Application Guide
5-3
8.
During the cycle, take several samples of the process material and
determine the average of the sample densities.
9.
CAL/STD ratio from latest CAL:
When the calibration cycle is
complete, the ratio of the last calibration measurement to the
standardization value is displayed.
10.
Use latest CAL value for CAL point 1:
To use the calibration value for
CAL point 1, enter
2
. Enter
1
if you do not want to use the value for
CAL point 1.
11.
Calibration is complete and the loop can be returned to automatic
control.