Thermo Scientific
4-1
4
Installation
Do not exceed the electrical and temperature ratings printed on the dataplate of
the control console.
Location
Keep line voltage variations to a minimum for best control accuracy. Do not locate
unit in areas of wide ambient temperature variation, such as near vents or outdoor
entrances. Allow at least three inches space all round the console, although more
space may be required for ease of maintenance.
Power Wiring
“A” voltage control consoles are designed to operate on 110V or 120V 50/60Hz,
“C” voltage consoles on a 208V or 240V 50/60Hz single phase power source. The
customer is to provide a main power disconnect switch or circuit breaker and
correctly sized power and ground wires according to local electrical codes. The
wiring should correspond with those carrying similar loads built into the console.
Improper operation of the control console could result in
dangerous conditions. To preclude hazard and minimize risk,
follow all instructions and operate within design limits noted on the
dataplate.
For personal safety and trouble-free operation, this unit must be
properly grounded before it is used. Always conform to the
National Electrical Code and Local Codes. Utilize proper
grounding techniques to reduce RFI and EMI for electronic gear.
Do not connect unit to already overloaded power lines; lower
voltage to unit will decrease power to the heating elements.
Содержание CC58114PBC
Страница 1: ......
Страница 6: ...List of Figures ii Thermo Scientific ...
Страница 12: ...Introduction 2 2 Thermo Scientific ...
Страница 14: ...Unpacking 3 2 Thermo Scientific ...
Страница 20: ...Installation Thermocouple Wiring Installation 4 6 Thermo Scientific ...
Страница 22: ...Initial Start up 5 2 Heratherm Thermo Scientific ...
Страница 46: ...Communication Option Addresses for Multiple Controllers 7 6 Thermo Scientific ...
Страница 50: ...Excess Temperature Option Excess Temperature Controller Operation 8 4 Thermo Scientific ...
Страница 57: ...Replacement Parts Thermo Scientific Heratherm 11 3 Figure 5 Wiring Diagram ...
Страница 58: ...Replacement Parts 11 4 Heratherm Thermo Scientific ...
Страница 59: ...Thermo Scientific 12 1 12 Error Codes This chapter is not applicable ...
Страница 60: ...Error Codes 12 2 Thermo Scientific ...
Страница 62: ...WEEE Compliance 13 2 Thermo Scientific ...
Страница 63: ...Thermo Scientific 14 1 14 14 Spare Parts and Accessories This chapter is not applicable ...
Страница 64: ...Spare Parts and Accessories 14 2 Thermo Scientific ...
Страница 66: ...Device Log 15 2 Thermo Scientific ...