Operation - UT150 Controller
Programming Operation: Entering a Program
Thermo Scientific
6-9
Entering Programming Mode
To access the programming menu:
1. Assure the indicators beside "RUN" and "L" on the controller display off. If
either indicator is extinguished, press and hold the "SET/ENT" button until the
display shows "mod", Select “rES" in the lower display with the "arrow" buttons.
Press and release the “SET/ENT” button once.
2. Press the "SET/ENT" button for 3 seconds to display "modE" in the upper
display and "rES" (Reset) in the lower display.
3. Press and release “SET/ENT” until "LoC" is displayed. Make sure the display
below "LoC" is "0" (zero). If it is not "0", use “DOWN ARROW” to select ''0'' and
press and release "SET/ENT” button to register the change to "0".
4. Press and release the "SET/ENT" button until "PrG" is displayed.
5. At "PrG" display, press the “UP ARROW” to make the lower display "1".
6. Press and release the "SET/ENT" button once to enter the programming
menu.
Entering program Parameters
The first display is the Start Set Point parameter, shown as "SSP" in the upper
display. The value assigned to SSP is usually the current room temperature, 25
°C.
On the next page is an illustration of the program profile and a table of the
parameters entered.
Basic Ramp and Dwell Parameters:
1. Use the arrow buttons to select "25" in the below "SSP" display, then press and
release the "SET/ENT" button twice to enter this new value for "SSP" is correct
and does not need to be changed, press and release the "SET/ENT" button
once to advance to the "StC" display.
2. Next is the Start Code parameter, shown as "StC" in the upper display. The
value assigned to StC is usually "1". This will instruct the program to start with
the current measured temperature. Start Setpoint Press the "SET/ENT" button
to advance to next display.
3. The next parameter, “SP1”, is the first setpoint value that is desired in the
chamber and is considered
a ramp segment. Select this target temperature
I
f the controller buttons are NOT pushed for 2 minutes, the
controller will return to the regular operator mode/menu.
Содержание CC58114PBC
Страница 1: ......
Страница 6: ...List of Figures ii Thermo Scientific ...
Страница 12: ...Introduction 2 2 Thermo Scientific ...
Страница 14: ...Unpacking 3 2 Thermo Scientific ...
Страница 20: ...Installation Thermocouple Wiring Installation 4 6 Thermo Scientific ...
Страница 22: ...Initial Start up 5 2 Heratherm Thermo Scientific ...
Страница 46: ...Communication Option Addresses for Multiple Controllers 7 6 Thermo Scientific ...
Страница 50: ...Excess Temperature Option Excess Temperature Controller Operation 8 4 Thermo Scientific ...
Страница 57: ...Replacement Parts Thermo Scientific Heratherm 11 3 Figure 5 Wiring Diagram ...
Страница 58: ...Replacement Parts 11 4 Heratherm Thermo Scientific ...
Страница 59: ...Thermo Scientific 12 1 12 Error Codes This chapter is not applicable ...
Страница 60: ...Error Codes 12 2 Thermo Scientific ...
Страница 62: ...WEEE Compliance 13 2 Thermo Scientific ...
Страница 63: ...Thermo Scientific 14 1 14 14 Spare Parts and Accessories This chapter is not applicable ...
Страница 64: ...Spare Parts and Accessories 14 2 Thermo Scientific ...
Страница 66: ...Device Log 15 2 Thermo Scientific ...