B·R·A·H·M·S GmbH
B·R·A·H·M·S KRYPTOR compact PLUS
User Manual (Version R08en)
Page
7
of 30
B·R·A·H·M·S GmbH, the protection provided by the
equipment may be impaired.
Biohazardous
When working with human serum, controls and cali-
brators all accessible parts of the analyzer must be
consi dered as biohazardous. The pipette tip, sample
cassettes, reagent cassettes, carousel drip pan, and
analyzer deck should be routinely disinfected.
Concerning all areas exposed to patient potential
infectious material or to user contact follow the clean -
ing requirements and use 5% Hypochloride solution
for decontamination. (See B.R.A.H.M.S KRYPTOR com-
pact PLUS Manual, chapter maintenance.)
It is strongly recommended to wear gloves and a
special coat.
Waste
The reagents and the waste must disposed of as
potential infectious laboratory waste according to local
regulations.
Dilution plate, reaction plate
Before disposing of the dilution plate as well as the
reaction plate stick the adhesive cover film with
bio hazard sign on the plate. The adhesive covers are
supplied with the plates.
B.R.A.H.M.S KRYPTOR compact SOLUTIONS 1 to 4
For detailed information read the safety instructions for
use inserts.
Safety data sheets are available at International Product
Support on request.
Laser
Concerning the sample carousel barcode reader (Class
2 laser) and the manual barcode reader do not look at
the laser. Avoid direct eye exposure.
Class 2
lasers are limited to a maximum output power
of 1 milliwatt and the beam must have a wavelength
be tween 400 and 700 nm. A person receiving an eye
exposure from a Class 2 laser beam, either accidentally
or as a result of someone else’s deliberate action (mis -
use) will be protected from injury by their own natural
aversion response. This is a natural involuntary res-
ponse that causes the individual to blink and avert their
head thereby terminating the eye exposure. Repeated,
deliberate exposure to the laser beam may not be safe.
The laser inside the instrument (Class 3B laser) emits
an invisible radiation whose characteristics are the fol-
lowing:
Laser Type
LTB
SRS
Beam Deviation (mrad)
3*3
5*8
Pulse Length (ns)
2.5
<3.5
Max Peak Power (kW)
100
45
Repetition Rate (Hz)
20
20
Class 3B
lasers may have sufficient power to cause an
eye injury, both from the direct beam and from reflec-
tions. The higher the output power of the device the
greater the risk of injury. Class 3B lasers are therefore
considered hazardous to the eye.
The laser is only accessible by opening the instrument
cover. The instrument cover (side access panel) is
equipp ed with a safety interlock, which stops laser
beam im me dia tely.
For further information see latest version of
B.R.A.H.M.S KRYPTOR compact PLUS Manual and
latest version of B.R.A.H.M.S KRYPTOR instruc-
tions for use for reagents and consumables.
Содержание BRAHMS KRYPTOR compact PLUS
Страница 1: ......
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 7: ......
Страница 8: ......
Страница 9: ......
Страница 10: ...B R A H M S KRYPTOR compact PLUS User Manual Version R08 106172 I SW 7 06 00 I Date 21 02 2018 ...
Страница 38: ......
Страница 40: ......
Страница 45: ......
Страница 55: ......
Страница 58: ......
Страница 59: ......
Страница 60: ......