Run controls
Control
Adding a Control through Worklist
- Select
Add control
in the
Data
,
Worklist
:
- Select a control from the list, press
Add to worklist,
Close, Close
.
Adding a Control to be tested on specific reagent
kit
- Double click on the kit which should be used to measure
the controls.
- Select
Control lot
from the drop-down arrow list.
- Select
Low volume
and/or
Stat status
for this control.
- Click on
Add to worklist, OK, Close
.
- Request a carousel hood opening by pressing on blue
button in front of instrument.
- Place the control tube on the sample carou-
sel making sure the barcode fills the slot.
- Close the carousel hood.
- Start the instrument:
Results
- Open results list:
- If result is in “needs resolution” status related to con-
centration out of 2SD range (shown in red) or with
flag, check the information and validate, reject or
relaunch the test. Accept ing the control automatically
sends the result to the QC program.
- Select the B.R.A.H.M.S KRYPTOR compact Quality
control program
to follow control life.
Run patient samples
with barcode
Patient Samples with barcode
- Select
Add sample
from
Data
,
Worklist
:
- Enter the Sample ID. Cassette ID and Tube position
are mandatory in case of sample without barcode.
- Select tests and dilutions from the displayed list,
Add
.
- Click on
Stat
to pipette a sample in priority.
- Click on
Low volume sample vial
if sample is in
microcup.
- Select
Save sample
and continue with required
samples,
Close
.
- Request a carousel hood opening by pressing on blue
button in front of carousel part.
- Place sample tubes on the carousel making
sure the barcode fills the slot.
- Close the carousel hood.
- Start the instrument:
Results
- Open results list:
- Select a test.
- It is possible to
Cancel
tests in detecting, or counting
step or in pending, only in “Ready/…” status.
- Results needing resolution can be:
Accepted
,
Re-run
or
Rejected
.
- Select the results then
Print Status
or
Print Report
to print results.
End of day
Shutdown (stand by modus)
- Check inventory of reagents on board for tomorrow’s
workload.
- Check the level of waste, distilled or demineralized
water and B.R.A.H.M.S KRYPTOR BUFFER.
- Check if the B.R.A.H.M.S KRYPTOR compact SOLU-
TION 3 and 4 is on board.
- Select
and confirm with
Yes
.
- Unload the reaction plate and confirm the popup
window with
OK
.
- Close the transparent damper and confirm the popup
window with
OK
.
To power OFF the instrument
- Remove dilution plate(s), samples, and reagents.
- Close carousel hood.
- Select
and confirm with
Yes
.
- Unload the reaction plate and confirm the popup win-
dow with
OK
.
- Close the transparent damper and confirm the popup
window with
OK
.
- Instrument is washing, wait for the completion of
steps (SOLUTION 3 and water washing) in case of
instrument power off.
- Close B.R.A.H.M.S KRYPTOR compact software:
- Close Windows by clicking on
Start
and
Turn off
.
- Switch OFF the PC, screen, printer and instrument.
B·R·A·H·M·S GmbH
B·R·A·H·M·S KRYPTOR compact PLUS
User Manual (Version R08en)
Page
20
of 30
Содержание BRAHMS KRYPTOR compact PLUS
Страница 1: ......
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 7: ......
Страница 8: ......
Страница 9: ......
Страница 10: ...B R A H M S KRYPTOR compact PLUS User Manual Version R08 106172 I SW 7 06 00 I Date 21 02 2018 ...
Страница 38: ......
Страница 40: ......
Страница 45: ......
Страница 55: ......
Страница 58: ......
Страница 59: ......
Страница 60: ......