Backup Log Files.
Backup of log of run and main -
tenance log files.
-
Remove messages from box office.
This action
enables the purge of the box office and avoid the
occurrence of data transmission problems.
-
Cleaning of water bottle.
Empty and replace dis tilled /
demineralized water if it is more than seven days old.
-
Clean carousel pan.
Clean the tray under the sample
carousel with a paper towel slightly wet with water,
then repeat with alcohol or specific product to decon-
taminate.
Monthly maintenance
-
Bottles decontamination.
Empty the bottles and
pour in 1 liter of 5% sodium hypochlorite solution. Fit
a bottle cap and swirl the liquid inside the bottle so
that it comes into contact with all the internal sur -
faces. Allow to stand for 15 minutes, then empty and
rinse the bottle with water.
-
Secure tip cleaning.
Tip parking over night or if instru-
ment is not in use for a longer time. B.R.A.H.M.S
KRYPTOR compact SOLUTION 3 and 4 must be ava il -
able.
-
Automatic check dot.
Tip coming on dot point. If
adjustment is correct, tip coming back to wash cup.
In case of adjustment problem, user should open the
fluidic hood and check if tip can be slightly bent to
enter the hole.
Reagent registration
For the following working steps installation of valid
version of K-DISK ANA is needed.
Necessary when using new lots only:
Reagent
- Click on
Reagent Lot Manager
:
- Click on
Register
.
- Scan the barcode sheet.
- Click on
OK
and confirm the registration with
Yes
.
Calibrator
- Click on
Calibrator/Standard Manager
:
- Click on
Register
.
- Scan the barcode sheet.
- Click on
OK
and confirm the registration with
Yes
.
Control
- Click on:
- Click on:
- Click on
Register new Controls
.
- Scan the barcode sheet.
- Click on
OK
.
Reconstitution of reagents
- Click on
to start reconstitution process.
- All freeze dried reagent kits placed in the instrument
will be reconstituted.
- In the status bar the progression of the reconstitution
in % is visible.
Perform a calibration
Calibrator
- Double click on the kit you want to calibrate.
- Select a calibrator lot from the drop-down arrow list.
- Click on
Add to worklist, OK, Close
.
- Request a carousel hood opening by pressing on blue
button in front of instrument.
- Place the calibrator tube on the sample car -
ousel making sure the barcode fills the slot.
- Close the carousel hood.
- Start the instrument:
Results
- Open results list:
- Select the calibrator result and click on
Validate curve
.
- The software indicates if the new calibration curve is
acceptable or not:
if Yes click on:
if No click on:
Содержание BRAHMS KRYPTOR compact PLUS
Страница 1: ......
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 7: ......
Страница 8: ......
Страница 9: ......
Страница 10: ...B R A H M S KRYPTOR compact PLUS User Manual Version R08 106172 I SW 7 06 00 I Date 21 02 2018 ...
Страница 38: ......
Страница 40: ......
Страница 45: ......
Страница 55: ......
Страница 58: ......
Страница 59: ......
Страница 60: ......