Thermo Scientific AquaSensors
™
AquaTrace
™
Trace DO System User Guide
24
▲
◄
►
MENU ESC ENTER
▼
3.
Determine the DO level of the current process sample using laboratory methods.
4.
Press the ENTER key to proceed with the calculation.
▲
◄
►
▼
102.8 ppb
Sensor ready?
MENU ESC ENTER
5.
The AV38 will display the current DO value for editing. Use the arrow keys to display the actual
value as determined by laboratory methods. When the value is correct, press the ENTER key to
continue.
1-point Sample?
( 102.8) ppb
▲
◄
►
▼
▲
◄
►
MENU ESC ENTER
▼
6.
The AV38 will update the span calibration in the sensor based on the current measurement.
During this time, the calibration procedure can be aborted by pressing the ESC key.
Zero
Calibrating…
MENU ESC ENTER
7.
If an error occurs during calibration that causes the procedure to fail, the reason for the failure
will be shown.
8.
After the calibration point has been stored, the AV38 will return to the calibration menu. The
user has the option of repeating the procedure if desired. If not, press the ESC key to exit the
menus and return to normal operations.
≡
Calibrate
►
1-Point Sample