7.
The AV38 will update the span calibration in the sensor per the calculated value for saturated air
using the recorded air pressure and measured temperature. During this time, the calibration
procedure can be aborted by pressing the ESC key.
Thermo Scientific AquaSensors
™
AquaTrace
™
Trace DO System User Guide
21
▲
◄
►
▼
Air
Calibrating…
MENU ESC ENTER
8.
If an error occurs during calibration that causes the procedure to fail, the reason for the failure
will be shown.
9.
After the calibration point has been stored, the AV38 will return to the calibration menu. The
user has the option of repeating the procedure if desired. If not, press the ESC key to exit the
menus and return to normal operations.
≡
Calibrate
►
Air
▲
◄
►
MENU ESC ENTER
▼
10.
Place the sensor back into the sample chamber and resume process flow through the
AquaTrace system.