Note
Dial numbers on front panel are for reference only; they are not
values for water temperature within the unit.
s
Warning
Do not leave the bath unattended during setting procedure.
s
1. Turn both the Temperature Control and High Temperature Limit
thermostat knobs fully clockwise. The Temperature Control Pilot lamp
should illuminate.
2. When the bath temperature reaches the desired temperature, slowly
turn the High Temperature Limit Thermostat counterclockwise and
stop when the High Temperature Limit Pilot Lamp illuminates.
3. Turn the High Temperature Limit Thermostat Knob clockwise to next
highest reference number. The High temperature Limit light should go
off. The High Temperature Limit thermostat is now set approximately
5° above the desired bath temperature.
4. Turn the Temperature Control Thermostat counterclockwise until
both pilot lamps are off.
5. Turn the Temperature Control thermostat clockwise until the
Temperature Control pilot lamp just lights. Allow the unit to stabilize,
re-adjust/fine-tune the Temperature Control thermostat as necessary.
5-4
GP Water Baths
Thermo Scientific
Section 5
Operation
Setting Thermostats