GP Water Baths
4-1
Thermo Scientific
Section 4
Installation
To obtain the most uniform operating conditions and results, place the
bath in an area remote from drafts, ventilation outlets, radiators, and other
rapidly changing ambient conditions.
To assure proper ventilation, allow at least 4" clearance between the rear,
top, and sides of the unit and adjacent wall. If two or more units are
positioned side-by-side, allow at least 8" between them.
Environmental Conditions
This instrument is designed to operate safely under the following
conditions:
• Indoor Use Only
• Temperature: 5° to 40°C
• Maximum Relative Humidity: 80% for temperatures to 22°C
• Maximum Altitude 2000 meters
Maximum performance is assured across the following temperature range:
• 15°C to 45°C