Rate of Deficit increase has exceeded
perforation alarm limit. Slow deficit increase
or disable alarm.
E
RROR
M
ESSAGES
A
Message Center
displays any Errors and
provides the user with instructions.
In the example to the right
indicates there
is no fluid detected by the bubble detector.
Fluid warming will be disabled.
Connect canister ring. If connected and
message persists call for support.
Indicates no volume change detected while
pump is running.
If problem persists, call for
support.
Open door detected, close door to resume,
DO NOT REMOVE CARTRIDGE. Do not
remove cartridge from unit during operation.
If removed and re-installed, unit may fail to
read actual fluid pressure accurately.
A new cartridge must be installed in device. If
a procedure has been started, use the end
procedure button to return to setup screens,
then press error message to dismiss. Proceed
through screens to re-prime new cartridge.
- or -
Stop vibrations.
If problem persists, call for
support.
- or -
No disposable cartridge detected. Insert
cartridge until click is felt. If cartridge
inserted and message
persists, call for
support.
Remove, discard, and replace tube set.
Suction clean out port on back of unit.
If the
problem persists, call for support.
Ensure tubing is fully inserted into bubble
detector. Check for bubbles in bubble
detector.
If problem persists, call for support.
Device requires re-calibration.
Call for
support.