37 |
P a g e
Part Number - 00824 Revision K, DCO 2017-0074
© 2017 Thermedx
®
, LLC
Section 11 – Specifications
Mechanical dimensions:
Height (on stand)
55 inches
Width:
26 inches (660mm)
Depth:
26 inches (660mm)
Weight:
95 pounds
Operation:
15.5°C to 25.5°C, 30 to 75% relative humidity
Transportation:
-40°C to 70°C, 25 to 95% relative humidity, and pressure within 50 to 106 kPa.
Storage:
0°C to 40°C, 25 to 95% relative humidity, and pressure within 50 to 106 kPa.
Temperature set-point: user selectable 37- 40°C
Tolerance: between initial fluid temperature and set-point +3°C
Over temperature protection: 46°C +3/-0°C
Irrigation bag capacity: 1-5 liters per hook (1 bag per hook).
Suction canister capacity: 3 liters maximum per canister. Up to four (4) suction canisters can be
accommodated.
Fluid measurement accuracy: +/- 250ml or +/- 10% of fluid pumped (whichever is greater).
This device is not manufactured using natural rubber latex components.
P2000: Electrical 120VAC, 60 Hz, 16A, (requires dedicated 20A outlet)
W2100: Electrical 240VAC, 50/60 Hz, 8A (requires dedicated 10A outlet)
Electrical Type, Protection Against Electrical Shock Class I Equipment, Type BF Applied Part
Mode of Operation Continuous
Type of Current Alternating
Ingress Protection Rating IPX1
MEDICAL EQUIPMENT
WITH RESPECT TO ELECTRIC SHOCK,
FIRE AND MECHANICAL HAZARDS ONLY
IN ACCORDANCE WITH UL 60601-1,
IEC/EN 60601-1, CAN/CSA C22.2 No. 601.1
47LG