Section 4 – Principle of Operation
The FluidSmart
®
uses a peristaltic pump to pressurize fluids from fluid bags hanging on
either side of the unit.
Fluid is conducted through a warming mechanism that uses infrared lamps to warm the
irrigation fluids on demand to a specified set temperature, generally approximating body
temperature. The warming mechanism may be enabled during approved procedures.
Flow rates are calculated by counting pump rotations and tracking changes in the weight
of fluid bags via load cells associated with the fluid bag hooks.
IR (infrared) temperature sensors are used to measure the fluid temperature in a non-
contact method. Three sensors are used in the warming mechanism (beginning, middle,
and end of fluid path). Two additional temperature monitors are used downstream of the
warming mechanism as a safety measure independent of software.
Using the flow rate, inlet temperature, and desired set-point temperature, the IR lamp
power is calculated. The warming process is adjusted as needed based on feedback from
temperature sensors.
Temperature measurements displayed by the system are readings of fluid temperature at
the exit of the warming mechanism. These readings do not account for temperature losses
as fluid passes through the disposable tubing.
Fluid volume instilled (inflow) and fluid deficit are determined on the basis of weight using
load cells in the fluid bag hooks and suction canister ring.
Fluid pressure is monitored via pressure transducers. Redundant sensors are used for safe
operation. Pressure measurements displayed by the system are readings at the system
and may not reflect the pressure at the surgical site.
The system operates via pressure control. The system will attempt to maintain the
specified pressure (i.e. 100mmHg). Pump speed will increase or decrease accordingly.
On the basis of the selected clinical discipline and procedure the system will enable
procedure specific pressure, flow, warming, and/or deficit monitoring alarm values.
o
The default values are user adjustable within specified safety limits. Not all
features are available for all procedures.