background image

12 

General Troubleshooting 

Problem 

Possible Cause 

Remedy 

The unit does not start after connection, tank 
filling, and pressing the On button 

Main fuses blown 

Replace blown fuses 

Motor defective 

Contact factory 

Reset tripped 

Reset 

Control circuit breaker tripped 

Reset 

Motor buzzes after pressing the On button 
and overload trips 

Voltage on two of the phases only 

Check incoming power supply, check and 
replace blown fuse 

Motor defective 

Contact factory 

No water circulation, even though the pump 
is rotating 

Pump is rotating in the wrong direction 

Change two of the incoming power leads 

Process water lines clogged 

Clean lines 

The unit does not heat 

Contactor defective 

Replace contactor 

Thermostat defective 

Replace thermostat 

Heating element defective 

Replace heating element 

Safety fuse defective 

Replace safety fuse 

The unit does not cool 

Solenoid valve at “Cooling Water In” defective  Replace solenoid valve 

Cooling heat exchanger clogged 

Clean coil 

Thermostat defective 

Replace thermostat 

The unit cooling all the time 

Dirt in the cooling solenoid valve 

Take apart the valve, clean out, replace if 
required 

Thermostat defective 

Replace thermostat 

Water comes out the overflow pipe 

Tank overfilled 

Drain some water 

Dirt in the fill solenoid valve 

Take apart the valve, clean out, replace if 
required 

Level sensing probe dirty or defective 

Clean probe, replace if required 

The unit does not fill the tank, level lamp is 
not on 

Level sensing probe dirty or defective 

Clean probe, replace if required 

Defective microprocessor 

Contact factory 

The unit does not fill the tank, level lamp is on 

Water supply not connected 

Connect supply 

Defective fill solenoid valve 

Replace if required 

 

Содержание Vactherm RV

Страница 1: ......

Страница 2: ......

Страница 3: ...Running Hour Counter 5 Displaying Clock Time 5 Setting Clock Time 5 Displaying Day of the Week 5 Setting Day of the Week 5 Timer On Off 5 Display Status of Timer 5 Turning Timer On Off 5 Setting Star...

Страница 4: ...t Temperature 10 Setting Set Point Temperature 10 Step 6 Adjusting the Vacuum 11 Preventive Maintenance 11 Once a Week 11 Once a Month 11 Once a Year 11 General Troubleshooting 12 Charts and Drawings...

Страница 5: ...eneral Mandatory Action Wear Eye Protection Wear Protective Gloves Wear Ear Protection Disconnect Before Carrying Out Maintenance or Repair Connect an Earth Terminal to Ground Only qualified personnel...

Страница 6: ...trols or any other major component torn loose from its mounting point Record any signs of concealed damage and file a shipping damage claim immediately with the shipping company Most carriers require...

Страница 7: ...percentage of imbalance is then 452 442 x 100 452 2 2 This exceeds the maximum allowable of 2 There is a terminal block for main power connection to the main power source The main power source should...

Страница 8: ...mperature Adjust button and the Password C button to initiate an automatic cool down cycle The controller will open the cooling valve and keep it open until the temperature of the water in the reservo...

Страница 9: ...our clock and will begin keeping dates from that point forward until reset To adjust the day press the Date Time button twice release and immediately use the Increase and Decrease buttons to set the c...

Страница 10: ...tons Once the desired day is displayed use the On button to toggle the status of the timer for that day Examples of days of week displays Monday timer off 1 Monday time on 1 Tuesday timer off 2 Tuesda...

Страница 11: ...rrent tank temperature in the display Password Entry Pressing the Password Entry buttons will allow for secured access to the control program and will prevent unauthorized personnel from making change...

Страница 12: ...t Circuit Temperature display flashing Alarm horn on and X1 output powered Temperature Sensor Disconnected Temperature display flashing Alarm horn on and X1 output powered Draining the Mold The temper...

Страница 13: ...and cumulative pressure rating of the maximum pump pressure and the unit fill fluid pressure NOTE A cooling source inlet filter ships loose in the crate with the unit Install this in the inlet of the...

Страница 14: ...F 7 2 C 5 40 F 4 4 C 10 35 F 1 7 C 15 30 F 1 1 C 20 25 F 3 9 C 25 20 F 6 7 C 30 CAUTION When your application requires the use of glycol use industrial grade glycol specifically designed for heat tran...

Страница 15: ...m the set point there may be a problem with the unit If this is the case refer to the General Troubleshooting section of this manual or contact our Customer Service Department 2 Check the pump dischar...

Страница 16: ...ent defective Replace heating element Safety fuse defective Replace safety fuse The unit does not cool Solenoid valve at Cooling Water In defective Replace solenoid valve Cooling heat exchanger clogge...

Страница 17: ...13 Charts and Drawings Figure 1 Pump Performance 60 Hz...

Страница 18: ...14 Figure 2 Cooling Capacity...

Страница 19: ...15 Figure 3 High Voltage Wiring Diagram...

Страница 20: ...16 Figure 4 Low Voltage Wiring Diagram...

Страница 21: ...17 Figure 5 Control Board Details...

Страница 22: ...18 Figure 6 Heating Element Details...

Страница 23: ...perature sensor 9 Microprocessor control 10 Heat exchanger 11 Heating element 12 Overflow pipe 13 Solenoid valve for water filling 14 Solenoid valve for cooling 15 Connection for external sensor 16 Ch...

Страница 24: ...20 Notes...

Страница 25: ...21 Notes...

Страница 26: ...22 Notes...

Страница 27: ......

Страница 28: ...Sales Customer Service Parts Department 847 966 2260 info thermalcare com 847 966 2260 847 966 9358 fax 847 966 2636 847 966 8560 847 966 9358 fax www thermalcare com info thermalcare com service the...

Отзывы: