THERA-Trainer TT-SENS-14-01 Скачать руководство пользователя страница 14

14

 

Art. no.: A003-928_Rev 01 

Last edited: CW40/2014 

 

 

FEEDBACK SENSOR INCL. TOUCH KEY 
USER MANUAL 

 

 

 Caution 

 Warning 

 

 

The feedback sensor incl. touch key is supplied with a USB charging cable as 
standard. This can be used to charge the battery from a USB interface (computer, 
control and display unit, etc.) without a feedback sensor USB charger.  

 

Do not use within 1.5 metres of the direct surrounding of the patient.  

 

► 

Only use USB chargers recommended by the manufacturer.  

 

Standard USB charger for feedback sensor:  

 

Available from the manufacturer as an option.  

-   Is not a medical-device USB charger for the feedback sensor.  
-   Do not use within 1.5 metres of the direct surrounding of the patient.  

 

Use a commercially-available USB charger (with CE marking).  

-   5 V (d.c.)/0.5 A 
-   USB connection  
-   Do not use within 1.5 metres of the direct surrounding of the patient.    

 

The service life of the battery depends on its charging/discharging cycles and 
storage temperature.  

 

Possible energy performance of batteries:  

 

Approx. 1,000 partial discharges 

 

The feedback sensor incl. touch key is fitted with over-discharge protection 

for the battery.  

 

When the battery is discharged:  

 

When the remaining charge (based on the capacity still available) is under 
30 %, the activation/charge status LED will flash green-orange.  

 

When the remaining charge (based on the capacity still available) is under 
15 %, the activation/charge status LED will flash orange.  

 

If the battery is fully discharged, the activation/charge status LED will be off.  

 

An operating period of approx. 10 hours per charging cycle can be obtained 
depending on the conditions of use.  
The battery capacity will decline over time. 

 

The manufacturer recommends having the battery replaced by a trained dealer 
after 2 to 3 years, depending on the number of charging cycles.

 

 
 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Faulty USB charging cable connection due to USB cable 
being plugged in incorrectly!

 

 

Make sure that the USB charging cable plug is correctly  

         positioned in the charging socket when it is connected.  

 

Observe position of plug.  

 

Only use a USB charging cable approved by the  

manufacturer.  

Using an unsuitable USB charging cable or USB charger is a 
fire hazard!

 

 

Only use a USB charging cable and USB charger for  

 

feedback sensor approved by the manufacturer.  

USB  Charging 
cable

 

 

 
 
 

 

USB charger

 

 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Operating 
period

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Charging 
requirements

 

 
 
 

 

Operating time 
 

 
 

Note

 

 
 
Charging 
process 
 
 

Содержание TT-SENS-14-01

Страница 1: ...FEEDBACK SENSOR INCL TOUCH KEY MODEL TT SENS 14 01 USER MANUAL English translation of original user manual...

Страница 2: ...et to know your feedback sensor incl touch key better It takes you safely through the functions and use of your new feedback sensor incl touch key and also offers plenty of tips and advice for getting...

Страница 3: ...h receiver for sensors 7 4 Switching on THERA Trainer software 8 OPERATION 11 8 1 Charging the battery 11 8 2 Switching on off the feedback sensor incl touch key 13 8 3 Establishing a connection 13 8...

Страница 4: ...on warns of the risk of material damage Warning warns of the risk of physical injury Danger warns of the risk of fatal injury Structure of warnings Type and source of danger Action to avoid the danger...

Страница 5: ...household waste Dispose of the product in accordance with local regulations Type BF medical device IP22 Protection against solid foreign objects with a diameter greater than 12 5 mm Protection against...

Страница 6: ...m range of movement must not overstrain the user Only use the feedback sensor incl touch key if in good and functional condition In case of damage malfunctions etc of the feedback sensor incl touch ke...

Страница 7: ...incl touch key Specialised staff Persons who gained their skills and expertise by means of professional training in the medical or medical technical sector e g doctors therapists specialist dealers Pe...

Страница 8: ...kits etc These can influence the device and should therefore be kept at a distance of 3 8 m 5 SCOPE OF DELIVERY AND EQUIPMENT The feedback sensor incl touch key comes with individual options This use...

Страница 9: ...for feedback sensor The feedback sensor incl touch key can only be used in combination with the suitable THERA Trainer software a control and display unit and a Bluetooth receiver for sensors 5 4 Remo...

Страница 10: ...IEW 1 Sensor orientation 2 Touch key with touch pad 3 Sensor detection here feedback sensor incl touch key Type A 4 Socket for USB charging cable 5 Feedback sensor incl touch key applied part Type BF...

Страница 11: ...to a THERA Trainer balo using the feedback sensor mounting kit for the THERA Trainer balo To fix it follow the instructions below Mount the sensor holder 1 to the system bracket 3 with two bolts 2 us...

Страница 12: ...er 4 to the system bracket 2 with two bolts 3 using a suitable tool Fix the sensor holder on the system bracket using the correct orientation wider side facing left Insert two bolts through the holes...

Страница 13: ...THERA Trainer software will be interrupted and can only be restored by restarting 7 4 Starting THERA Trainer software Switch on control and display unit THERA Trainer software starts automatically Wh...

Страница 14: ...discharge protection for the battery When the battery is discharged When the remaining charge based on the capacity still available is under 30 the activation charge status LED will flash green orang...

Страница 15: ...a medical device USB charging cable and USB charger Charging must take place outside a 1 5 metre radius of the direct surrounding of the patient Charging process begins The LED of the charge activatio...

Страница 16: ...der to establish a wireless connection a Bluetooth receiver for the sensors must be plugged into the control and display unit Only use the Bluetooth receiver for sensors approved by the manufacturer a...

Страница 17: ...indicator flashes red The control and display unit shows the error code Observe the user manual for the relevant THERA Trainer software The software version of the feedback sensor incl touch key is r...

Страница 18: ...dback sensor incl touch key must always be disconnected from the mains before cleaning and disinfecting Make sure that no cleaning agent or disinfectant gets into the feedback sensor incl touch key If...

Страница 19: ...h key is suitable for further use by a different customer Before the feedback sensor incl touch key is reused Clean and disinfect thoroughly see chapter 10 Cleaning and disinfection Perform a visual c...

Страница 20: ...tive poles have been taped up Feedback sensor incl touch key Dimensions l x w x h approx 7 5 cm x 5 5 cm x 2 5 cm Weight approx 80 g Protection category IP22 only if the feedback sensor incl touch key...

Страница 21: ...alue at 10 m distance AC mains power supply Wired interference 0 15 0 5 66 dB uV to 56 dB uV quasi peak value CISPR 11 56 dB uV to 46 dB uV mean value The limit values decrease linearly with the logar...

Страница 22: ...0 4 11 0 5 period test voltage level 5 Ut 5 periods test voltage level 40 Ut 25 periods test voltage level 70 Ut A Voltage interruption 5 duration 5 s C Fast transients EN 61000 4 4 2 kV A Surge volta...

Страница 23: ...ienced radio TV technician for help 14 5 Statement acc RSS Gen Issue 3 Sect 7 1 3 This device complies with Industry Canada licence exempt RSS standard s Operation is subject to the following two cond...

Страница 24: ...ntechnik GmbH Blumenweg 8 88454 Hochdorf Germany Phone 49 7355 93 14 0 Fax 49 7355 93 14 15 E Mail info thera trainer de Internet www thera trainer de Article no A003 928_Rev 01 Last edited CW40 2014...

Отзывы: