FEEDBACK SENSOR INCL. TOUCH KEY
USER MANUAL
Caution
Warning
The feedback sensor incl. touch key is supplied with a USB charging cable as
standard. This can be used to charge the battery from a USB interface (computer,
control and display unit, etc.) without a feedback sensor USB charger.
Do not use within 1.5 metres of the direct surrounding of the patient.
►
Only use USB chargers recommended by the manufacturer.
Standard USB charger for feedback sensor:
Available from the manufacturer as an option.
- Is not a medical-device USB charger for the feedback sensor.
- Do not use within 1.5 metres of the direct surrounding of the patient.
Use a commercially-available USB charger (with CE marking).
- 5 V (d.c.)/0.5 A
- USB connection
- Do not use within 1.5 metres of the direct surrounding of the patient.
The service life of the battery depends on its charging/discharging cycles and
storage temperature.
Possible energy performance of batteries:
Approx. 1,000 partial discharges
The feedback sensor incl. touch key is fitted with over-discharge protection
for the battery.
When the battery is discharged:
When the remaining charge (based on the capacity still available) is under
30 %, the activation/charge status LED will flash green-orange.
When the remaining charge (based on the capacity still available) is under
15 %, the activation/charge status LED will flash orange.
If the battery is fully discharged, the activation/charge status LED will be off.
An operating period of approx. 10 hours per charging cycle can be obtained
depending on the conditions of use.
The battery capacity will decline over time.
The manufacturer recommends having the battery replaced by a trained dealer
after 2 to 3 years, depending on the number of charging cycles.
Faulty USB charging cable connection due to USB cable
being plugged in incorrectly!
Make sure that the USB charging cable plug is correctly
positioned in the charging socket when it is connected.
Observe position of plug.
Only use a USB charging cable approved by the
manufacturer.
Using an unsuitable USB charging cable or USB charger is a
fire hazard!
Only use a USB charging cable and USB charger for
feedback sensor approved by the manufacturer.
USB Charging
cable
USB charger
Operating
period
Charging
requirements
Operating time
Note
Charging