![The Armstrong Monitoring Corporation AMC-400 Series Скачать руководство пользователя страница 53](http://html1.mh-extra.com/html/the-armstrong-monitoring-corporation/amc-400-series/amc-400-series_user-manual_1099482053.webp)
AMC-400 Series Digital Transmitter with Smart Sensor 2
49
Data Register
MODBUS
Register
Index
Monitor
Access
Serial Number; a 16bit number between 1-
65535 exclusive to a specific AMC-400 model.
0
*1 Read
Hardware Rev / Software Rev; a 16bit
number made up of four bit fields:
5 bit hardware version [15:11] (1-32
versions)
3 bit hardware revision [10:8] (A-H
revisions)
5 bit software version [7:3] (1-32 versions)
3 bit software revision [2:0] (A-H revisions).
These fields are displayed on the LCD during
power up.
1
*1 Read
Transmitter Status / Mode; a 16bit number
made up of two 8 bit fields representing
transmitter status and modes.
Instrument Status Bits[15:8]:
Transmitter Okay =0x00
Transmitter Error
=0xFF
Instrument Mode Bits[7:0]: (Default 0x8F)
Reserved for Self Test=1
(always being self-tested)
Bit7
Reserved for Remote
Calibration Type=0
(Remote Calibration not
supported)
Bit6
Reserved for Output
Enable=0
(Output not available)
Bit5
Reserved for Output
Type=0
(Output not applicable)
Bit4
Reserved for Baud Rate=1
(Fixed at 9600)
Bit3
Reserved bit=1
Bit2
Reserved bit=1
Bit1
Reserved bit=1
Bit0
2
*1 Read