5.3.1 REQUIRED
EQUIPMENT; EXCHANGE PROGRAM
For qualified personnel, the following is a recommended list of equipment required:
5/16 nut driver and Phillips and Robertson screw driver set
TOOL SCREWDRVR SLOT ESD 1.5MM, Wiha 27215.
To replace Smart Sensor 2 Module with a pre-calibrated unit, first completely remove power from Digital
Transmitter by unplugging Power and Communication Connector as shown in. Use a nut driver to remove
nut that is holding Smart Sensor 2 Module to Digital Transmitter. Carefully remove and replace Smart
Sensor 2 Module attached to SS2 Female Connector. Replace nut holding Smart Sensor 2 to Digital
Transmitter with nut driver.
Warning:
It is possible to improperly insert SS2 Male Pins into SS2 Female connector. Ensure
pins are lined up properly with these connectors before inserting.
5.3.2 REQUIRED EQUIPMENT; ON-SITE CALIBRATION
For qualified personnel, the following is a recommended list of calibration equipment required.
TOOL SCREWDRVR SLOT ESD 1.5MM, Wiha 27215
Calibration kit and Cal Gas, please refer to Section 2.1.2.
5.3.3 CALIBRATION
PROCEDURE
The AMC-400 Digital Transmitter is equipped with a calibration feature allowing for one-man
calibration at the transmitter location. In most cases it is only Zero and Span adjustments that
are made in field calibrations. All Calibration procedures will automatically exit to normal
operation without change if left inactive for several minutes. Please refer to Appendix for menu
flow diagrams and guide
.
Note:
During calibration new gas concentration is not updated to monitor and last gas
concentration value will be reported.
Refer to Figure 5-1 Front Panel User Interface to perform the following calibration procedure:
1) The zero calibration can perform a calibration with any background air where the
monitored gas is greater than 0 PPM (as example 5ppm CO gas could be used
instead of 0ppm). Specifics are described in 3
rd
step of this process. Perform
gassing with a calibration adapter that is firmly inserted into front panel of AMC-400
Digital Transmitter; please refer to 2.1.2 SMART SENSOR 2 MODULE(S)
SPECIFICATION for gas flow rate and Calibration kit.