7.
Open the clamp on the patient line.
8.3.3 Procedure
1.
Push the standby/on button to put the device in on mode.
2.
Open the slider a few millimetres until the set point indicator LED is off.
3.
Flush the hot water into the standard set.
4.
Make sure that the over temperature alarm starts.
5.
Close the clamps.