15
2-11-2. Remote sensing
CAUTION
!
When using shielded sense wires, ground the shield in one place only.
The location can be the power supply chassis or one of the output terminals
Use remote sense where the load regulation at the load end is critical. In remote sense, the power supply will
compensate for voltage drop on the load wires.
◆
Refer to
for the maximum voltage drop on load wires.
The voltage drop is subtracted from the total voltage available at the output. Follow the instructions below to configure
the power supply for remote sensing:
1. Ensure that the AC On/Off is in the Off position.
2. Remove the local sense jumpers from J2.
3. Connect the negative sense lead to terminal J2-5 (-S) and the positive sense lead to terminal J2-1(+S) of the
J2 mating connector. Ensure that the J2 mating connector is plugged securely into the rear panel sense
connector, J2.
4. Turn On the power supply.
CAUTION
!
1. If the power supply is operating in remote sense and either the positive or negative
load wire is not connected, an internal protection circuit will activate and shut
down the power supply. To resume operation, turn the AC On/Off to the Off
position, connect the open load wire, and turn On the power supply.
2. If the power supply is operated without the remote sense lines or local sense
jumpers, it will continue to work, but the output voltage regulation will be degraded.
Also, the OVP circuit may activate and shut down the power supply.
2-11-3. J2 sense connector technical information
- J2 connector type
:MC1.5/5-G-3.81, Phoenix.
- Plug type
:MC1.5/5-ST-3.81, Phoenix.
- Wire AWG
: 28 up to 16.
- Stripping length
: 7mm.
- Tightening torque
: 0.22Nm to 0.25Nm ( 2.2kgf.cm to 2.6kgf
・
cm )
WARNING
!
There is a potential shock hazard at the sense point when using power supply with a
rated output voltage greater than 40V. Ensure that the connections at the load end
are shielded to prevent accidental contact with hazardous voltages.