25
4-6. Output shut-off (SO) control via rear panel J1 connector
Contacts 2, 3 and 15 of J1 (Fig.3-2, Item 24) serve as Output Shut-Off (SO) terminals. The SO terminals accept a
2.5V to 15V signal or Open-Short contact to shut off the power supply output. To rise the output again after shut-off,
apply the reverse-logic signal (Table 4-2) to the SO terminals. This function is useful for connecting power supplies in
a “Daisy-chain”. The SO control can be used also to reset the OVP and Fold Protection.
◆
Refer to
"4-2. Over voltage protection (OVP)"
and
"4-3. Under voltage limit (UVL)"
for details.
When the unit is shut-off by J1 signal,
the VOLTAGE display will show “SO” to indicate the unit state. J1 contact 15 is
the SO signal input and contacts 2 and 3, IF_COM, are the signal return (connected internally). Contacts 2, 3 and 15
are optically isolated from the power supply output. The SO control logic can be selected by the rear panel SW1
Setup switch.
◆
Refer to Table 4-2 for SW1 setting and SO control logic.
CAUTION
!
In Auto-Restart mode the output will re-rise automatically after reset the SO control,
but in safe start mode pushing OUTPUT button is needed to raise the output again
after reset
Table 4-2: SO logic selection
SW1-5 setting
SO signal level
J1-2(3), 15
Power supply output
Display
Down (default)
2V to 15V or Open
0V to 0.6V or Short
On
Off
Voltage/Current
“SO”
Up
2V to 15V or Open
0V to 0.6V or Short
Off
On
“SO”
Voltage/Current
This function cannot be used with Output OFF. Please turn ON an output using the function of output
interception (SO).
4-7. Enable/disable control via rear panel J1 connector
Contacts 1 and 14 of J1 (Fig.3-3, Item 24) serve as Output Enable/Disable terminals by switch or relay. This function is
enabled or disabled by the SW1 Setup switch position 9.
◆
Refer to Table 4-3 for Enable/Disable function andSW1setting.
Table 4-3: Enable/Disable function and SW1 setting
SW1-9 setting
Enable/Disable inputs
Power supply output
Display
ALARM LED
Down (default)
Open or Short
On
Voltage/Current
Off
Up
Open
Off
“ENA”
Blinking
Short
On
Voltage/Current
Off
CAUTION
!
To prevent possible damage to the unit, do not connect any of the Enable/Disable
inputs to the positive or negative output potential.
If the Enable/Disable inputs are opened in Safe Start mode, it is required to short the Enable /Disable inputs
and then press OUTPUT button or send OUT1 command to resume operation. But in auto-restart mode set
these inputs short, and then the output will rise up automatically after output-off by this function.