48
6-7-2. Service Request: Enable and Event Registers
The conditional Registers are continuously monitored. When a change is detected in a register bit which is enabled,
the power supply will generate an SRQ message.
The SRQ message is: "!nn" terminated by CR, where the nn is the power supply address.
The SRQ will be generated either in Local or Remote mode.
◆
Refer to Tables 6-10 to 6-13 for details of the Enable and Event registers.
1. Fault Enable Register
The Fault Enable Register is set to the enable faults SRQs.
Table 6-10: Fault Enable Register
BIT
Enable bit name
Fault symbol
Bit Set condition
Bit Reset condition
0(LSB)
Spare bit
SPARE
User command:
“FENA nn” where nn is
hexadecimal
User command:
“FENA nn” where nn is hexadecimal
(if nn=”00”, no fault SRQs will be
generated).
1
AC Fail
AC
2
Over Temperature
OTP
3
Foldbck
FOLD
4
Over Voltage
OVP
5
Shut Off
SO
6
Output Off
OFF
7(MSB)
Enable
ENA
2. Fault Event Register
The Fault Event will set a bit if a condition occurs and it is enabled. The register is cleared when FEVE? , CLS
or RST commands are received.
Table 6-11: Fault Event Register
BIT
Enable bit name
Fault symbol
Bit Set condition
Bit reset condition
0(LSB)
Spare bit
SPARE
Fault condition
occurs and it is enabled.
The fault can set a bit,
but when the fault clears
the bit remains set.
Entire Event Register is cleared when
user sends “FEVE?” command to
read the register.
“CLS” and power-up also clear the Fault
Event Register.
1
AC Fail
AC
2
Over Temperature
OTP
3
Foldbck
FOLD
4
Over Voltage
OVP
5
Shut Off
SO
6
Output Off
OFF
7(MSB)
Enable
ENA
3. Status Enable register
The Status Enable Register is set by the user to enable SRQs from changes in power supply status.
Table 6-12: Status Enable Register
BIT
Status name
Status symbol
Bit Set condition
Bit reset condition
0(LSB)
Constant Voltage
CV
User command:
“SENA nn” is received,
where nn is hexadecimal
bits.
User command:
“SENA nn” is received, where nn is
Hexadecimal bits.
If “nn”=00, no SRQ is sent where there
is a change in Status Condition Register.
1
Constant Current
CC
2
Not Fault
NFLT
3
Fault active
FLT
4
Auto-Restart
enabled
AST
Always zero
Always zero
5
Fold enabled
FDE
Always zero
Always zero
6
Spare
SPARE
Always zero
Always zero
7(MSB)
Local Mode
LCL
“SENA nn” command
“SENA nn” command