27
LRR
Time between:
Source 1 first rising edge and
Source 2 last rising edge.
LFR
Time between:
Source 1 first rising edge and
Source 2 last falling edge.
LRF
Time between:
Source 1 first falling edge
and Source 2 last rising
edge.
LFF
Time between:
Source 1 first falling edge
and Source 2 last falling
edge.
Phase
The phase difference of two
signals, calculated in
degrees.
360
x
t2
t1
Note
The in-built help system can be used to see
detailed automatic measurement definitions.
2-2-2. Add Measurement
The
Add Measurement
function allows you to add up to eight
automatic measurement items on the bottom of the screen from any
channel source.
Add
Measurement
Item
1. Press the
Measure
key.
Measure
2. Press
Add Measurement
from the
bottom menu.
3. Choose either a
V/I
,
Time
or
Delay
measurement from the side menu
and choose the type of
measurement you wish to add.