104
Trigger On
–
Data
If Data or Address/Data was configured for the
Trigger On setting, then the number of bytes,
data and addressing mode (I
2
C) can be
configured.
1. Press
Data
from the bottom menu.
2. Press
Number of Bytes
from the
side menu and choose the number
of bytes for the data.
I
2
C
1~5 Bytes
3. Press
Addressing Mode
to toggle
between 7 and 10 bit addressing
modes.
4. Press
Data
from the side menu to
edit the triggering data.
To edit the data, use the
Variable
knob to highlight a binary or hex
digit and press
Select
. Use the
Variable
knob to choose a value for
the digit and press
Select
to
confirm.
VARIABLE
Binary
0,1,X (don’t care)
Hex
0~F, X (don’t care)
Trigger On -
Address
If Address or Address/Data was configured for
the Trigger On setting, then the triggering
address must be configured.
1. Press
Address
on the bottom
menu.
2. Press
Addressing Mode
to toggle
between 7 and 10 bit addressing
modes.
3. To choose a preset address as the
default address, press
Choose
Preset
and select a preset
address.
Address
Description
0000 000 0
General Call
0000 000 1
START Byte
0000 1XX X
Hs-mode
1010 XXX X EEPROM
0000 001 X
CBUS