Overview

www.ti.com

1

Overview

This document is intended to serve as a user’s guide for the ADS5XJ6x EVM. The ADS5XJ6XEVM is an
evaluation module (EVM) designed to evaluate the ADS58J63 and ADS54J66 high-speed, JESD204B
interface ADCs. The EVM includes an onboard clocking solution (LMK04828), transformer coupled inputs,
full power solution, and easy-to-use software GUI and USB interface.

The following features apply to this EVM:

Transformer-coupled signal input network allowing a single-ended signal source from 0.25 MHz to 750
MHz

LMK04828 system clock generator that generates FPGA reference clocks for the high-speed serial
interface and may be used to generate the ADC sampling clock (default setting)

Transformer-coupled clock input network to test the ADC performance with a very low-noise clock
source

High-speed serial data output over a standard FMC interface connector

The ADS5XJ6XEVM is designed to work seamlessly with the TSW14J56EVM, Texas Instruments’
JESD204B data capture/ pattern generator card, through the High Speed Data Converter Pro (HSDC Pro)
software tool for high-speed data converter evaluation. The ADS5XJ6XEVM was also designed to work
with many of the development kits from leading FPGA vendors that contain an FMC connector.

1.1

Required Hardware

The following equipment is

included

in the EVM evaluation kit.

ADS5XJ6X Evaluation Board (EVM)

+5-V DC power cable

Mini-USB cable

The following list of equipment are items that are

not included

in the EVM evaluation kit but are items

required for evaluation of this product in order to achieve the best performance:

TSW14J56EVM Data Capture Board, two 5-V power supplies, and Mini-USB cable

+5-V DC power supply with a minimum current source of 3 Amps

Computer running Microsoft

®

Windows

®

8, 7, or XP

One low-noise signal generator. Recommendations:

RF generator, > +17 dBm, < –40 dBc harmonics, < 500 fs jitter 20 kHz–20 MHz, 10 MHz–2 GHz
frequency range

Examples: TSW2170EVM, HP HP8644B, Rohde & Schwarz SMA100A

Band-pass filter for desired analog input. Recommendations:

Band-pass filter,

60-dB harmonic attenuation,

5% bandwidth, > +18-dBm power, < 5-dB

insertion loss

Examples: Trilithic 5VH-series Tunable BPF, K&L BT-series Tunable BPF, TTE KC6 or KC7-series
Fixed BPF

Signal path cables, SMA and/or BNC with BNC-to-SMA adapters

2

ADS5XJ6X Evaluation Module

SLAU641D – June 2015 – Revised January 2016

Submit Documentation Feedback

Copyright © 2015–2016, Texas Instruments Incorporated

Содержание ADS5XJ6XEVM

Страница 1: ...tup 5 3 ADS54Jxx GUI Low Level View Tab 6 4 HSDC Pro GUI Main Panel 7 5 Channel 1 Data Capture Results from Quick Start Procedure 8 6 Channel 1 Data Capture Results in Bypass Mode 9 7 ADS54Jxx GUI 12...

Страница 2: ...uation The ADS5XJ6XEVM was also designed to work with many of the development kits from leading FPGA vendors that contain an FMC connector 1 1 Required Hardware The following equipment is included in...

Страница 3: ...es the locations of the EVM features Figure 1 EVM Feature Locations 1 4 References The following links provide convenient references ADS5XJ6XEVM software schematics layout BOM available at www ti com...

Страница 4: ...ADS5XJ6XEVM It must be used to properly configure the devices on the EVM 1 Download the ADS54Jxx GUI from the TI website The References section of this document contains link to find the software on...

Страница 5: ...The rest of this section will describe the hardware setup steps Figure 2 Quick Start Test Setup 2 2 1 TSW14J56EVM Setup First setup the TSW14J56EVM with following the instructions 1 Connect the ADS54J...

Страница 6: ...is lit in the top right corner of the GUI If it is not lit click the Reconnect USB button and check the USB Status indicator again If it is still not lit then verify the EVM is connected to the compu...

Страница 7: ...6 This will take about 3 seconds 7 Enter 245 76M into the ADC Output Data Rate field at the bottom left corner then click outside this box or press return on the PC keyboard 8 The GUI will display the...

Страница 8: ...y In the ADS54Jxx GUI click on the Low Level View tab then click the Load Config button Navigate to C Program Files 86 Texas Instruments ADS54Jxx GUI Configuration Files select the file called LMK_Con...

Страница 9: ...ration files Click OK Click the Instrument Options menu at the top of HSDC Pro and select Reset Board Make sure the display mode is set to Real FFT Click Capture in HSDC Pro to capture data from the A...

Страница 10: ...Serial Device is recognized when the EVM is connected to the PC Verify that the green USB Status LED light in the top right corner of the GUI is lit If it is not lit press the Reconnect FTDI button Tr...

Страница 11: ...d better SNR results The clock and analog inputs must be frequency locked such as through 10 MHz references in order to achieve coherency 3 3 HSDC Pro Settings HSDC Pro has some settings that can help...

Страница 12: ...l View tab to load configuration files LMK04828 Enables control of many of the LMK04828 features Configuration files can be used to setup the LMK04828 in known working configurations however this tab...

Страница 13: ...n in the Register Map with the values shown in the Register Map log The log can be viewed by double left clicking in the bottom left status bar of this page Read Register Button Read from the register...

Страница 14: ...o provide the device clock to the TSW14J56 and the SYSREF signals to both boards This option will provide the best performance as long as the clock source has better phase noise performance than the L...

Страница 15: ...CXO The LMK04828 is used as a clock generator using the onboard 122 88 MHz VCXO JP2 must be shorted to turn on the onboard VCXO The internal PLLs of the LMK04828 can be used with the onboard VCXO to g...

Страница 16: ...tion described in Table 8 Table 8 Connector Description Connector Description J1 Channel A analog input J17 Channel B analog input J18 Channel C analog input J5 Channel D analog input J6 External ADC...

Страница 17: ...ecember 2015 to D Revision Page Changed first bullet in the not included items list in the Required Hardware section to two 5 V power supplies 2 Changed Quick Start Test Setup image 5 Changed content...

Страница 18: ...ing the warranty period to the address designated by TI and that are determined by TI not to conform to such warranty If TI elects to repair or replace such EVM TI shall have a reasonable time to repa...

Страница 19: ...transmitter has been approved by Industry Canada to operate with the antenna types listed in the user guide with the maximum permissible gain and required antenna impedance for each antenna type indic...

Страница 20: ...ified allowable ranges some circuit components may have elevated case temperatures These components include but are not limited to linear regulators switching transistors pass transistors current sens...

Страница 21: ...REMOVAL OR REINSTALLATION ANCILLARY COSTS TO THE PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES RETESTING OUTSIDE COMPUTER TIME LABOR COSTS LOSS OF GOODWILL LOSS OF PROFITS LOSS OF SAVINGS LOSS OF USE L...

Страница 22: ...sponsible for compliance with all legal regulatory and safety related requirements concerning its products and any use of TI components in its applications notwithstanding any applications related inf...

Отзывы: