18
TT-SX 900x Series High-Voltage Differential Probes User Manual
6 Specifications & Characteristics
The probe and oscilloscope should be warmed up for at least 20 minutes before
any testing and the environmental conditions should
not exceed the probe’s
specified limits.
HINWEIS
Alle Angaben in den folgenden Tabellen sind Kennwerte, sofern
nicht anders angegeben.
NOTE
All entries included in the following tables are characteristics
unless otherwise stated.
REMARQUE
Toutes les entrées incluses dans les tableaux suivants sont des
caractéristiques, sauf indication contraire.
Safety Specifications
IEC/EN 61010-031:2015
Measurement Category III
6.1
Electrical Specifications
Typical data as well as nominal and measured values are not warranted by Testec.
TT-SX 9001
TT-SX 9002
Part Number
15601
15602
Bandwidth (-3 dB)
30 MHz
30 MHz
Gain Accuracy
±2%
±2%
Attenuation Ratio
10x / 100x
20x / 200x
Rise Time
14 ns
14 ns
Absolute Maximum Rated
Input Voltage
(each side to ground)
1000 V CAT III
1000 V CAT III
Maximum Differential Input
Voltage (DC + Peak AC)
±70 V at 10x
±700 V at 100x
±140 V at 20x
±1400 V at 200x
Maximum Common Mode Input
Voltage (DC + Peak AC)
±70 V at 10x
±700 V at 100x
±140 V at 20x
±1400 V at 200x
Input Impedance
5
MΩ, 2 pF
(each side to ground)
5 MΩ, 2 pF
(each side to ground)
Output Voltage
±7 V
(driving 1
MΩ load)
±7 V
(driving 1
MΩ load)