3. Maintenance
24
Input VSWR (unloaded) Specification: < 2.5 800 MHz ~ 1 GHz, 1.5 GHz ~ 6 GHz
1) Set the Network Analyzer as follows:
Start: 30 kHz
Stop: 6000 MHz
Measure: S11
Format: SWR
2) Calibrate the Network Analyzer in order to measure S11.
3) Connect the calibrated cable to the input port of TC-5062C and measure SWR.
4) Verify S11 SWR of-5062C meets its specification.
3.2.4 Typical VSWR of TC-5062C
VSWR (30 KHz to 6000 MHz), Scale 1/Div
[Figure 8] Typical VSWR of TC-5062C