3. Maintenance
20
3.1 Maintenance of TC-5062C
The TC-5062C is designed and built to last for a long life and with minimal and easy
maintenance.
Optimal RF shielding is obtained using a shield form gasket between the case and door. It
must be checked periodically for any damage or excessive wear that would compromise
the seal. Pressure on the gasket by the door results in a proper RF seal.
NOTE
Preventive maintenance requires you check the RF gasket regularly for excessive
wear. Also check the gasket cavity for any black oxide build-up.
3.1.1 Cleaning
Periodically wipe the case with a damp cloth mild detergent; do not use abrasives or
solvents.
3.1.2 Check List for Maintenance
1) Check for loose screws and tighten with proper tools, if necessary.
2) Check for a loose connection. If a loose connector is found, tighten the
connector with proper tools.
3) Check for a damaged cable, especially near the connector-cable neck. Replace
any damaged cables found.
4) Visually inspect the RF seal between the TEM Cell and door for excessive wear.
A worn gasket will not have as much height protruding from the cavity as a new
or good gasket.
A worn gasket may exhibit a frayed mesh. Replace the gasket if excessive wear
is observed.