21.  GENERAL SPECIFICATION 

Waveform 

Asymmetrical rectangular 

Amplitude  

(over 500 Ohm load) 

99.0 mA  

+/- 10% 

Max intensity 

50V zero to peak 

Constant voltage over 470-1500 Ohm 

Constant current over 160-470 Ohm 

Output plug 

Fully shielded 

Channels 

Single channel 

Batteries 

2 x AA alkaline (two AA batteries) or 2 x AA NiMH 

Weight 

75 gms

 

without batteries 

Dimensions 

120 x 60 x 20 mm 

Safety Classification 

Internal power source. 

Environmental Specifications:  

Operating:     

     

                   

Storage: 

Temperature range: 5 to 40

C  

Humidity: 15 to 93% RH non-condensing 

Atmospheric pressure: 700hPa to 1060hPa 

Temperature range: -25-+70

C  

Humidity: Up to 93% RH non-condensing  

Atmospheric pressure: 700hPa to 1060hPa 

 TYPE BF  

EQUIPMENT 

Equipment  providing  a  degree  of  protection  against 
electric shock, with isolated applied part. 

Designed for continuous use. 

 

This symbol on the unit means “Refer to Instructions for 
Use” 

 IP22 

The unit is not water resistant, and should be protected 
from liquids. 

 

Complies with EU WEEE regulations 

Applied Part 

Vaginal  and  anal  electrodes.  Optional  skin  surface 
electrode pads. See section 18. 

Contact duration: At least 10 minutes. 

 

Note:

 The electrical specifications are nominal and subject to variation from 

the listed values due to normal production tolerances of at least 5%. 

Содержание Perfect PFE

Страница 1: ...0 INSTRUCTIONS FOR USE READ CAREFULLY BEFORE USE...

Страница 2: ...ert the vaginal probe Press and hold the ON button to switch the device on Regulate the output intensity with the buttons and Press and hold the OFF button to switch the device off ON Button P UP DOWN...

Страница 3: ...y tested products for the applications in the areas of gentle electrotherapy muscle toning continence management and pain relief during labour Please read these instructions for use carefully and keep...

Страница 4: ...GRAMMES 15 11 CONTENT 16 12 UNIT INFORMATION 17 12 1 CONTROLS DISPLAY 17 12 2 OPERATING INSTRUCTIONS 18 13 SETTING UP AND USING THE PERFECT PFE 20 13 1 INSTALLATION OF BATTERIES 20 13 2 CONNECTING LEA...

Страница 5: ...ue number so that the device can be identified This medical device is not water resistant and should be protected from liquids Atmospheric Pressure indicates the atmospheric limits to which the medica...

Страница 6: ...ectly complements pelvic floor exercises The perfect PFE is very easy to use with four clearly labelled preset training programmes a fifth programme for chronic pelvic pain and a simple push button co...

Страница 7: ...h to zero and flashes LEADS if the connection comes loose Automatic Keypad Lock Automatic keypad lock prevents any accidental changes in settings Large LCD Screen Clearly shows the operation of the un...

Страница 8: ...as a bulge in the vagina or a feeling of heaviness discomfort pulling dragging or dropping This occurs when one or more of the pelvic organs bladder bowel or uterus become displaced and sag down into...

Страница 9: ...g the muscles you identified earlier don t tense the muscles in your legs stomach or buttocks and remember to breathe normally 4 Aim to hold each squeeze or contraction for three to five seconds then...

Страница 10: ...in medical rehabilitation and training in competitive sports EMS produces intensive and effective muscular contraction In rehabilitation EMS is a well established method for treatment of a broad field...

Страница 11: ...ion medium frequencies 20 50 Hz can put a high level of strain on the muscle thus promoting the muscular structure Each preset programme has a specific frequency and pulse width that will offer the be...

Страница 12: ...sis see Ova cystitis sports injuries Raynaud disease fibromyalgia and neuralgia plantar fasciitis post operative pain and sometimes non painful conditions such as tiredness insomnia travel sickness ti...

Страница 13: ...ofessional before use Do NOT use when driving operating machinery or similar actions needing muscular control Loose electrode pads damaged leads or sudden changes in contact may cause brief involuntar...

Страница 14: ...unless the control unit is powered OFF as insertion or removal when stimulation is active may cause discomfort or tissue irritation Caution If tissue irritation occurs discontinue treatment immediate...

Страница 15: ...ay affect the sockets or cause the battery connector to develop a bad contact Caution Temperature Relative Humidity of storage 20 C 40 C 8 70 R H Temperature Relative Humidity of transportation 20 C 4...

Страница 16: ...t starts becoming apparent after about four weeks The sensation is like a strong drawing in of the muscles of the vagina pulling up the pelvic floor Your natural reaction will be to pull your muscles...

Страница 17: ...ng the TONE programme is a mixture of a strong drawing in of the muscles and then releasing A strong and fit pelvic floor may increase sexual health and enjoyment PAIN Shown on the screen as PAIN The...

Страница 18: ...TION 12 1 CONTROLS DISPLAY Programme selected Programme selector Time selector OFF button ON button Keypad Lock Programme duration min Strength selected mA Strength up Strength down Low Battery Output...

Страница 19: ...unit will remain in the WORK part of the cycle for 5 seconds while intensity is being adjusted The strength levels are shown on the LCD The strength control buttons will not operate until the unit is...

Страница 20: ...unlock press and hold the strength down control button LEADS ALARM The perfect PFE monitors the connection and the contact between you and the probe or the pads This is to prevent sudden changes if a...

Страница 21: ...nit is unlikely to be used for a long period Some types of batteries may leak corrosive fluid Battery Life Batteries should last about 18 hours at full power Unused batteries have a nominal shelf life...

Страница 22: ...rientation makes no difference and you can ignore this colour coding 13 3 PREPARING FOR SESSION 1 Before using perfect PFE you will need to visit the toilet 2 Lubricate the metal electrode surfaces an...

Страница 23: ...ed P in the centre of the control unit keypad 3 With the required programme selected you can adjust the intensity of the muscle stimulation until you reach a comfortable level Once you have reached a...

Страница 24: ...strength display on the control unit will reduce to 0 and flash during the rest period The URGE and PAIN programmes work differently There is no need to have a contraction The strength should be comfo...

Страница 25: ...be restricted to one muscle group and the mucosal tissue has different electrical characteristics anal stimulation is less comfortable than vaginal You should consult your healthcare professional bef...

Страница 26: ...d otherwise by a healthcare professional 5 Anal probes with long electrodes the metal part that run up and down the length of the attachment should always be inserted with the metal parts facing hip t...

Страница 27: ...label from our website www tenscare co uk attach this to an envelope or padded bag with the unit enclosed and post it back to us Upon receipt we will process your old device for components recovery a...

Страница 28: ...e from manufacturing defects for 2 years from date of purchase This warranty is void if the product is modified or altered is subject to misuse or abuse damaged in transit lack of responsible care is...

Страница 29: ...chase receipt Please ensure all relevant details are completed before sending your unit in for service Please ensure your contact details are still current and include a brief description of the probl...

Страница 30: ...ill also prevent anyone from increasing the intensity to a high level without firm contact between the machine and the skin A connection error can occur if 1 A break has developed within one of the tw...

Страница 31: ...appen in childbirth or some surgical procedures Please consult your healthcare professional No sensation on one side of the probe or electrode Position is not optimal needs adjusting The current flows...

Страница 32: ...g Atmospheric pressure 700hPa to 1060hPa Temperature range 25 70 C Humidity Up to 93 RH non condensing Atmospheric pressure 700hPa to 1060hPa TYPE BF EQUIPMENT Equipment providing a degree of protecti...

Страница 33: ...ANTY NAME ADDRESS POSTCODE DAYTIME TELEPHONE E MAIL MODEL DATE OF PURCHASE ATTACH PROOF OF PURCHASE DO NOT SEND IN PROBE OR ELECTRODE PADS RETAILERS NAME RETAILERS ADDRESS RETAILERS POSTCODE BRIEF DES...

Страница 34: ...33 NOTES...

Страница 35: ...e limited https plus google com TenscareLtdEpsom https uk pinterest com TensCareLtd https tenscareltd wordpress com https www youtube com channel UCzpik9dmLlJ3j0aHOpQ 0sg EC Declaration of Conformity...

Страница 36: ...35 Distributed by Manufactured by TensCare Ltd 9 Blenheim Road Epsom Surrey KT19 9BE UK Tel 44 0 1372 723434 www tenscare co uk Pub No I PPFE UK Version 2 2 05 17...

Отзывы: