Tenma 72-6851 & 72-6853 Instruction manual
DELTAV? Sequential
command
Operation complete message generated immediately after <rmt> is sent.
Returns the delta voltage value in Volts in <nr2> numeric format. The syntax of the
response is
DELTAV <nr2><rmt>
Example:
If the delta voltage is 0·55 Volts the response to the command
DELTAV? will be DELTAV 0.55<rmt>
DELTAI? Sequential
command
Operation complete message generated immediately after <rmt> is sent.
Returns the delta current value in Amps in <nr2> numeric format. The syntax of the
response is
DELTAI <nr2><rmt>
Example:
If the delta current is 0·550 Amps the response to the command
DELTAI? will be DELTAI 0.550<rmt>
INCVV Sequential
command
Operation complete message generated immediately after execution.
Increment the output voltage by the delta voltage value and verify that the voltage is within
±3 counts or 5% of the target value. If the value of the output voltage is outside the range
for the specified output the value is set to the maximum value allowed, no error is
generated.
INCV Sequential
command
Operation complete message generated immediately after execution.
This command is identical to INCVV above except that no verification of the output voltage
is performed, thus saving the possible 500ms required to read back the output voltage.
This is useful when it is known that the slew time will be short or during constant current
operation.
DECVV Sequential
command
Operation complete message generated immediately after execution.
Decrement the output voltage by the delta voltage value and verify that the voltage is
within ±3 counts or 5% of the target value. If the value of the output voltage is outside the
range for the specified output the value is set to the minimum value allowed, no error is
generated.
DECV Sequential
command
Operation complete message generated immediately after execution.
This command is identical to DECVV above except that no verification of the output
voltage is performed, thus saving the possible 500ms required to read back the output
voltage. This is useful when it is known that the slew time will be short or during constant
current operation.
INCI Sequential
command
Operation complete message generated immediately after execution.
Increment the set current by the delta current value. If the value of the set current is
outside the range for the specified output the value is set to the maximum value allowed,
no error is generated.