Tenma 72-6851 & 72-6853 Instruction manual
∗
SRE <nrf>
Sequential command.
Operation complete message generated immediately after execution.
Set the Service Request Enable Register to <nrf>. If the value of <nrf>, after rounding,
is less than 0 or greater than 255 an execution error is generated and error number 119
(Value out of range) is placed in the Execution Error Register.
∗
SRE?
Sequential command.
Operation complete message generated immediately after <rmt> is sent.
Returns the value of the Service Request Enable Register in <nr1> numeric format. The
syntax of the response is
<nr1><rmt>
Example.
If the Service Request Enable Register contains 01000001b the
response
to
∗
SRE? will be 65<rmt>.
∗
STB?
Sequential command.
Operation complete message generated immediately after <rmt> is sent.
Returns the value of the Status Byte Register in <nr1> numeric format. The syntax of
the response is
<nr1><rmt>
Example.
If the Status Byte Register contains 01000001b the response to
∗
STB?
will be 65<rmt>.
∗
SAV <nrf>
Sequential command.
Operation complete message generated immediately after execution.
Save settings to store number <nrf>. The instrument settings are saved in the store
pointed to by <nrf>. If the value of <nrf>, after rounding, is less than 1 or greater than 25
an execution error is generated and error number 115 (illegal store number) is placed in
the Execution Error Register.
The scope of the
∗
SAV command is as follows:
The output is set to stored values of Volts, current, OVP, delta V and delta A.
The output status is set to the stored value.
∗
TST?
Sequential command.
Operation complete message generated immediately after <rmt> is sent.
Examine the state of the FLT bit in the Status byte register and return the result. If the
bit is clear the response is
0<rmt>
any failure is indicated by a non-zero FLT bit and the response will be
1<rmt>
The value in the FLT bit is 1 if an execution error 002 has been reported since the
instrument was powered up.
∗
WAI
Sequential command.
Operation complete message generated immediately after execution.
Wait for operation complete true. As all commands are completely executed before the
next is started this command takes no additional action.
24
www.tenma.com
electronic components distributor