Tenma 72-6851 & 72-6853 Instruction manual
∗
OPC
Sequential command.
Operation complete message generated immediately after execution.
Sets the Operation Complete bit (bit 0) in the Standard Event Status Register.
This will happen immediately the command is executed because of the sequential
nature of all operations.
∗
OPC?
Sequential command.
Operation complete message generated immediately after <rmt> is sent.
Query operation complete status. The syntax of the response is
1<rmt>
The response will be available immediately the command is executed because of the
sequential nature of all operations.
∗
PRE <nrf>
Sequential command.
Operation complete message generated immediately after execution.
Set the Parallel Poll Enable Register to the value <nrf>. If the value of <nrf>, after
rounding, is less than 0 or greater than 255 an execution error is generated and error
number 119 (Value out of range) is placed in the Execution Error Register.
∗
PRE?
Sequential command.
Operation complete message generated immediately after <rmt> is sent.
Returns the value in the Parallel Poll Enable Register in <nr1> numeric format. The
syntax of the response is
<nr1><rmt>
Example.
If the Parallel Poll Enable Register contains 01000001b the response to
∗
PRE? will be 65<rmt>.
∗
RST
Sequential command.
Operation complete message generated immediately after execution.
Resets the instrument as follows. The output is set to minimum voltage, minimum
current, maximum OVP, meter damping off and output OFF. No other action taken.
∗
RCL <nrf>
Sequential command.
Operation complete message generated immediately after execution.
Recall settings from store number <nrf>. The settings in the store pointed to by <nrf>
are recalled and installed as the instrument settings. If the value of <nrf> are recalled
and installed as the instrument settings. If the value of <nrf>, after rounding, is less than
1 or greater than 25 an execution error is generated and error number 115 (illegal store
number) is placed in the Execution Error Register. If the requested store is empty
execution error 116 (Recall empty store requested) is generated. If the contents of the
selected store are corrupt execution error 117 (Stored data is corrupt) is generated. Any
of the above error conditions will result in no change of the instrument settings.
The scope of the
∗
RCL command is as follows:
The output is set to stored values of Volts, current, OVP, delta V and delta A.
The output status is set to the stored value.