8
•
Then press
ENTER
to go into the tolerance setup and the “TOL±1%” icon will flash
on the LCD.
•
Press ◄/► to adjust the tolerance and then press
ENTER
to confirm, to allow you
to begin the sorting measurement.
•
Press
SORTING
to exit sorting measurement and return to normal mode.
CALIBRATION FUNCTION
•
Calibration function can effectively reduce stray impedance caused by test wires.
•
The meter offers short and open calibration. Short calibration is to remove the
impact of contact resistance and test wire resistance that interferes with low
impedance measurement.
•
Open calibration is to remove stray capacitance and resistance of test wires that
interfere with high impedance measurement.
Access Calibration Function
•
Power on the meter and press and hold
CAL
to
access open calibration.
•
“OPEn” will show on the secondary display (Fig. 1),
then press
CAL
again to begin the calibration and
the meter will count down from 30 seconds on the
LCD.
•
When it reaches zero “PASS” appears (Fig. 2),
which indicates open calibration has finished.
•
Press the
CAL
button and “Srt” appears on the
secondary display (see Fig. 3).
•
Insert short-circuit device into input terminals, press
CAL
to begin and the meter will count down from 30
seconds to 0 seconds.
•
“PASS” appears indicating that short calibration is
complete (Fig. 4).
•
Press
CAL
again to return to normal measurement
mode.
Note: if the the LCD shows “FAIL” under open
calibration mode, it indicates that the calibration has
failed. Check whether the input terminals are open
or not. Make sure the circuit is open and do the
calibration again.
•
Similarly, under short calibration mode if the LCD
shows “FAIL” it indicates that the calibration has
failed. Check the short circuit device has been
inserted into the input terminals.
•
Make sure the terminals are shorted and calibrate again.
PC COMMUNICATION
•
Insert the disc supplied and install software from auto run.
•
Press the
USB
button to access communication function.
•
Connect the meter to the PC using the USB wire and run the meter software on
the PC to start data transfer.
•
Press
USB
again to exit and stop the data transfer.
DPEn
Cal APO
PRSS
DPEn
Cal APO
Src
Cal APO
PRSS
Src
Cal APO
FRIL
DPEn
Fig. 1
Fig.
2
Fig.
3
Fig.
4
Fig.
5