13
10kHz
20.000Ω
0.001Ω
± (1.0% + 5)
200.00Ω
0.01Ω
± (1.0% + 5)
2.0000kΩ
0.1Ω
± (0.3% + 5)
20.000kΩ
1Ω
± (0.5% + 5)
200.00kΩ
0.01kΩ
± (1.0% + 5)
2.0000MΩ
0.1kΩ
± (2.0% + 5)
100kHz
20.000Ω
0.001Ω
± (2.0% + 5)
200.00Ω
0.01Ω
± (2.0% + 5)
2.0000k
0.1Ω
± (2.0% + 5)
20.000kΩ
1Ω
Function
: Resistance gear
Measurement mode
: DCR
Frequency
Range
Min.
Resolution
Accuracy:
±
(a% of r b digits)
(18°C to 28°C)
200.00Ω
0.01Ω
± (1.0% + 5)
2.0000kΩ
0.1Ω
± (0.3% + 5)
20.000kΩ
1Ω
± (0.3% + 5)
200.00kΩ
0.01kΩ
± (0.5% + 5)
2.0000MΩ
0.1kΩ
± (1.0% + 5)
20.000MΩ
1kΩ
± (2.0% + 5)
200.00MΩ
0.1MΩ
± (2.0% + 5)
Note: The precision is evaluated if D is less than 0.1.
Ae = Ae* √1+D²
if D exceeds 0.1.
(Ae: Precision).
BATTERY REPLACEMENT
•
Replace the battery as soon as the low battery symbol appears on the LCD, in
order to avoid an impact on measurement accuracy.
•
Replace the old battery with a new battery of the same specifi
c
ation (9V).
Содержание 72-10465