REL% MEASUREMENT

REL% mode measures % deviation between two components.

The measured value of the tested object shows on the main display, with the %

value on the secondary display.

Set the main display value as the nominal reference.

% display range: -99%~99.9%.

% calculation: REL%=(Dᶜᵘʳ-Dʳᵉᶠ)/Dʳᵉᶠ* 100%.

Dᶜᵘʳ: main display value of tested object.

Dʳᵉᶠ: the reference value that has been set.

If Dᶜᵘʳ > 2Dʳᵉᶠ or 2Dᶜᵘʳ < Dʳᵉᶠ, “OL%” will display on the secondary display and the

main display will show the measured value of the tested object.

Access REL mode

Press 

FUNC

 to select the desired “AUTO L”, “AUTO C”, “AUTO R” or “AUTO

DCR” mode.

Connect the tested object to the input terminals and press 

REL

 to access REL%

mode.

appears on the LCD with the main display showing the measured component

value and the secondary display showing % of deviation.

Press 

REL

 again and the reference value will show on the main display and the

icon flashes, while the % value will still be visible on the secondary display.

Press 

REL

 again to return to normal REL% measurement mode.

Press and hold 

REL

 to exit REL% measurement and return to normal mode.

SORTING MEASUREMENT

Sorting mode is used to quickly select components’ value within a specified

tolerance.

Press 

FUNC

 to select the desired “AUTO L”, “AUTO C”, “AUTO R” or “AUTO

DCR” mode.

Connect the tested object to the input terminals and press and hold 

FREQ

 to

access sorting mode. “Sorting” will appear on the LCD. This sets the limit value.

“PASS” shows on the main display and the measured component value on the

secondary display is set to a nominal value.

Connect another component and “PASS” will show if the component falls within the

set limit and the measured value shows on the secondary display.

The buzzer will sound once. If it is out of scope, “FAIL” shows and the measured

component value.

Press and hold 

FREQ

 to exit.

Set up sorting tolerance

The sorting limit can be set at the following values:

±0.25%, ±0.5%, ±1%, ±2%, ±5% ±10%, ±20%, +80%~-20%.

The meter defaults to ±1%. To set up sorting, put the meter under sorting mode

first and then press 

SETUP

 and the “Range” icon will flash on the LCD.

Press 

ENTER

 to confirm and set the main display parameter, of which the last digit

flashes.

Press ▼/▲ to decrease/increase the digit or ◄/► to select another flashing digit

for adjustment.

Содержание 72-10465

Страница 1: ...1 LCR METER Model 72 10465 ...

Страница 2: ... Mode 6 Measurement Frequency 7 REL Measurement 7 Sorting Measurement 8 Calibration Function 8 PC Communication 9 Backlight 9 Auto Power Off 9 Selection of Serial Parallel Mode 9 Inductance Measurement 9 Capacitance Measurement 10 Resistance Measurement 10 DCR Measurement 10 PC Communication Protocol 10 Technical Indicators 13 Battery Replacement 14 Maintenance WHAT S INCLUDED One LCR meter One us...

Страница 3: ...wered If the USB port is connected to the PC the meter is USB powered and will communicate with the PC for data collection MEASUREMENT RANGE ACCURACY L 20mH 2000H best accuracy 0 5 5 C 200pF 20µF best accuracy 0 5 5 R 20Ω 200MΩ best accuracy 0 3 5 Impedance Frequency DCR 100 120Hz 1kHz 10kHz 100kHz 0 1 1 1 0 1 0 1 0 1 0 1 0 1 10 0 5 0 5 0 5 0 5 0 5 10 100k 0 3 0 3 0 3 0 5 0 3 100k 1M 0 5 0 5 0 5 1...

Страница 4: ... Resistance and Reactance Reactance varies in proportion to the frequency of the AC circuit Impedance forms an imaginary vector comprised of Resistance R real and Reactance X imaginary with Impedance Z R jX which can also be represented by amplitude Z and phase angle θ which can be seen in the diagram If the phase angle θ 0 the reactance is inductive in nature if θ 0 the reactance is capacitive in...

Страница 5: ...surement mode for capacitance 6 Rᵖ Parallel measurement mode for resistance 7 Rˢ Serial measurement mode for resistance PRODUCT OVERVIEW 8 DCR DC resistance measurement mode 9 D Dissipation factor 10 Q Quality factor 11 θ Phase angle 12 ESR Equivalent serial resistance 13 EPR Equivalent parallel resistance 14 DUT Device under test 15 Short press 1 second long press 2 seconds indicating the time to...

Страница 6: ...METER UNDER L C R MODE Select the corresponding parameters under manual L C R mode Selection of main parameter Default status is AUTO LCR once powered on Select FUNC to select parameters of AUTO LCR AUTO L AUTO C AUTO R DCR AUTO LCR Selection of secondary parameter After a main parameter has been selected press SER PAL to switch between serial and parallel mode Press D Q θ to select D Q θ ESR ESR ...

Страница 7: ... mode Press and hold REL to exit REL measurement and return to normal mode SORTING MEASUREMENT Sorting mode is used to quickly select components value within a specified tolerance Press FUNC to select the desired AUTO L AUTO C AUTO R or AUTO DCR mode Connect the tested object to the input terminals and press and hold FREQ to access sorting mode Sorting will appear on the LCD This sets the limit va...

Страница 8: ... calibration has finished Press the CAL button and Srt appears on the secondary display see Fig 3 Insert short circuit device into input terminals press CAL to begin and the meter will count down from 30 seconds to 0 seconds PASS appears indicating that short calibration is complete Fig 4 Press CAL again to return to normal measurement mode Note if the the LCD shows FAIL under open calibration mod...

Страница 9: ... MEASUREMENT Press the power button and then press FUNC to display Lp on the LCD Select inductance measurement range Insert the inductor into input terminals or connect the inductor to the meter using the test clamp Press FREQ to select suitable testing frequency Press D Q θ to select the auxiliary parameter to measure CAPACITANCE MEASUREMENT Capacitors must be completely discharged before measure...

Страница 10: ...he power button Press the FUNC button to display DCR on the LCD and select DCR measurement range Insert the resistor into the input terminals or connect the resistor to the meter using the test clamp Note The secondary parameter and the frequency is negligible under DCR measurement and doesn t show on the LCD PC COMMUNICATION SPECIFICATION Communication parameters 1 Baud rate 9600 2 Data bit 8 3 S...

Страница 11: ... 0 5 200 00µH 0 01µH 1 0 5 2000 0µH 0 1µH 0 5 5 20 00mH 1µH 0 5 5 100kHz 20 000µH 0 001µH 2 0 5 200 00µH 0 01µH 2 0 5 2000 0µH 0 1µH 2 0 5 Function Inductance gear Measurement Mode Lˢ Lᵖ Note The precision is evaluated if D is less than 0 1 Ae Ae 1 D if D exceeds 0 1 Ae Precision Function Capacitance gear Measurement mode Cˢ Cᵖ Frequency Range Min Resolution Accuracy a of reading b digits 18 C to ...

Страница 12: ...nF 5 0 5 Note The precision is evaluated if D is less than 0 1 Ae Ae 1 D if D exceeds 0 1 Ae Precision Function Resistance gear Measurement mode Rˢ Rᵖ Frequency Range Min Resolution Accuracy a of reading b digits 18 C to 28 C 100Hz 120Hz 200 00Ω 0 01Ω 1 0 5 2 0000kΩ 0 1Ω 0 3 5 20 000kΩ 1Ω 0 3 5 200 00kΩ 0 01kΩ 0 5 5 2 0000MΩ 0 1kΩ 1 0 5 20 000MΩ 1kΩ 2 0 5 1kHz 20 000Ω 0 001Ω 1 0 5 200 00Ω 0 01Ω 1 ...

Страница 13: ...uracy a of reading b digits 18 C to 28 C 200 00Ω 0 01Ω 1 0 5 2 0000kΩ 0 1Ω 0 3 5 20 000kΩ 1Ω 0 3 5 200 00kΩ 0 01kΩ 0 5 5 2 0000MΩ 0 1kΩ 1 0 5 20 000MΩ 1kΩ 2 0 5 200 00MΩ 0 1MΩ 2 0 5 Note The precision is evaluated if D is less than 0 1 Ae Ae 1 D if D exceeds 0 1 Ae Precision BATTERY REPLACEMENT Replace the battery as soon as the low battery symbol appears on the LCD in order to avoid an impact on ...

Страница 14: ...ct the battery external power and power input terminal If the measurement does not work normally check if the test wires are in good condition and if the component pin has been connected with the input terminals Refer all servicing to qualified personnel Do not disassemble the meter replace components or change the circuits randomly INFORMATION ON WASTE DISPOSAL FOR CONSUMERS OF ELECTRICAL ELECTRO...

Отзывы: