3
IMPORTANT SAFETY INFORMATION
•
When using electrical appliances basic safety precautions should always be
followed.
•
Do not use the meter in environments exposed to explosive gas, vapour or dust, in
direct sunlight or high radiation.
•
There are no user serviceable parts in this product. Refer all servicing to qualified
personnel.
•
Turn off all the circuit power and discharge completely all capacitors before
measuring in-line components.
•
Measurement terminals, capacitors and other live components must be discharged
before being measured.
•
There are two power supplies available for the meter: 9V battery and USB
powered.
•
If the USB port is connected to the PC, the meter is USB-powered and will
communicate with the PC for data collection.
MEASUREMENT RANGE & ACCURACY
L: 20mH - 2000H best accuracy: (0.5% + 5)
C: 200pF - 20
µ
F best accuracy: (0.5% + 5)
R: 20Ω - 200MΩ best accuracy: (0.3% + 5)
Impedance/
Frequency
DCR
100/120Hz
1kHz
10kHz
100kHz
0.1-1
1.0%
1.0%
1.0%
1.0%
1.0%
1-10
0.5%
0.5%
0.5%
0.5%
0.5%
10-100k
0.3%
0.3%
0.3%
0.5%
0.3%
100k-1M
0.5%
0.5%
0.5%
1.0%
1M-20M
1.0%
1.0%
1.0%
20M-200M
2.0%
2.0%
5.0%
Remark
D ≤ 0.1
Note: Multiply by
√1+D²
if D exceeds 0.1
Formula to convert capacitance to impedance: Zᶜ = 1/2πfC
Formula to convert inductance to impedance: Zᶫ = 2πfL
AMBIENT CONDITIONS
•
Altitude: <2000 metres
•
Storage humidity: ≤ 75% RH
•
Operating environment: 0°C~40°C
•
Storage environment: -20°C~50°C
Please read these instructions carefully before use and retain for future
reference.
Содержание 72-10465