11
Frequency
Range
Min.
Resolution
Accuracy:
±
(a% of r b digits)
(18°C to 28°C)
100Hz /
120Hz
20.000mH
1µH
± (1.0% + 5)
200.00mH
0.01mH
± (0.5% + 5)
2000.0mH
0.1mH
± (0.5% + 5)
20.000H
1mH
± (0.5% + 5)
200.00H
0.01H
± (1.0% + 5)
2000.0H
0.1H
± (1.0% + 5)
1kHz
2000.0µH
0.1µH
± (1.0% + 5)
20.000mH
1µH
± (0.5% + 5)
200.00mH
0.01mH
± (0.5% + 5)
2000.0mH
0.1mH
± (1.0% + 5)
20.000H
1mH
± (1.0% + 5)
200.00H
0.01H
± (2.0% + 5)
10kHz
20.000µH
0.01µH
± (1.0% + 5)
200.00µH
0.01µH
± (1.0% + 5)
2000.0µH
0.1µH
± (0.5% + 5)
20.00mH
1µH
± (0.5% + 5)
100kHz
20.000µH
0.001µH
± (2.0% + 5)
200.00µH
0.01µH
± (2.0% + 5)
2000.0µH
0.1µH
± (2.0% + 5)
Function
: Inductance gear
Measurement Mode
: Lˢ/Lᵖ
Note: The precision is evaluated if D is less than 0.1.
Ae=Ae* √1+D²
if D exceeds 0.1.
(Ae: Precision).
Function
: Capacitance gear
Measurement mode
: Cˢ/Cᵖ
Frequency
Range
Min.
Resolution
Accuracy:
±
(a% of r b digits)
(18°C to 28°C)
100Hz /
120Hz
20.000nF
1pF
± (2.0% + 5)
200.00nF
0.01nF
± (0.5% + 5)
2000.0nF
0.1nF
± (0.5% + 5)
20.000µF
1nF
± (0.5% + 5)
200.00µF
0.01µF
± (1.0% + 5)
2000.0µF
0.1µF
± (2.0% + 5)
20.00mF
0.1mF
± (2.0% + 5)
Содержание 72-10465
Страница 1: ...1 LCR METER Model 72 10465 ...