www.element14.com
www.farnell.com
www.newark.com
www.cpc.co.uk
TM
TM
Page <8>
V1.0
22/08/18
Important Notice :
This data sheet and its contents (the “Information”) belong to the members of the Premier Farnell group of companies (the “Group”) or are licensed to it. No licence is granted
for the use of it other than for information purposes in connection with the products to which it relates. No licence of any intellectual property rights is granted. The Information is subject to change
without notice and replaces all data sheets previously supplied. The Information supplied is believed to be accurate but the Group assumes no responsibility for its accuracy or completeness, any
error in or omission from it or for any use made of it. Users of this data sheet should check for themselves the Information and the suitability of the products for their purpose and not make any
assumptions based on information included or omitted. Liability for loss or damage resulting from any reliance on the Information or use of it (including liability resulting from negligence or where the
Group was aware of the possibility of such loss or damage arising) is excluded. This will not operate to limit or restrict the Group’s liability for death or personal injury resulting from its negligence.
Tenma is the registered trademark of the Group. © Premier Farnell Limited 2016.
Operating Steps:
1. Turn the power in “off” situation, then remove the test lead out of the meter.
2. Use screwdriver to take off the screw on the battery cover. Then you can take off old battery for replacing
3. User screwdriver to take off two pieces of screws, then you can replacing the burned fuse(s) as replacement.
Fuse Specification
: F1 Fuse Ø6.35 × 31.8mm, F 10A H 600V
F2 Fuse Ø6.35 × 31.8mm, F 0.5A H 600V