2. DC Current Measurements (see diagram 3)

1. Turn rotary switch to “A 

” current measurement.

2.  Insert the red test lead into the “mA” or “10A” terminal, and the black test lead into the “COM” terminal, Connect the test leads 

across with the object being measured. The measured value shows on the display.

Warning:

 

•   

Pre-requisites: Turn off power to the circuit before the connections between the test leads across with the object 

being measured.

•   Selecting the correct terminal input and turn the rotary switch to select the measuring function. In case of no any 

idea on the value input of the current, just simply test from the high value to low one.

•  Fuses are located on mA and 10A terminal input.

•  Never attempt the test lead connect to any circuits especially on the power supply terminal and may be hurt.

•   For the safety purpose, less than 10 seconds is for each measurement duration and keep 15 minutes duration for 

next measurement during the current measurement over 5A

3. Resistance Measurements (see diagram 4)

1. Turn rotary switch to “Ω“ohm measurement.
2.  Insert the red test lead into the “Ω” terminal, and the black test lead into the “COM” terminal, Connect the test leads across 

with the object being measured. The measured value shows on the display.

 
•   

The LCD displays “OL” indicating open-circuit for the tested resistor or the resistor value is higher than the 

maximum range of the meter.

•   To maintain the resistance measurement accuracy, discount circuit power and discharge all the high voltage 

capacitors during the measuring resistance.

• 

Thetestleadscause0.1Ω~0.2Ωresistancevariationduringthemeasurement,inordertoobtainprecisionreading

in low-resistance measurement. Need to make the short circuit on the test leads and mark the measurement value 

which show on LCD display. Then deduct this variation value on the measurement value which come the meter, 

measurement value obtained from LCD display-Variation value on test leads = The actual measuring value.

• 

IfΩreadingwithshortenedtestleadsisnotlessthanorequalto0.5Ω,checkforloosetestleads,incorrectfunctions

election or others.

• 

Forhigh-resistancemeasurementgreaterthan1MΩ,itisnormaltotakeseveralsecondstoobtainastablereading

with short test leads for measurements.

•  Do no input the higher than DC 60V and AC 30V voltage to prevent any damage and accident.

Содержание 72-10390A

Страница 1: ...www element14 com www farnell com www newark com www cpc co uk TM TM Page 1 V1 0 22 08 18 Palm Size Digital Multimeter 72 10390A User Manual ...

Страница 2: ...rating manual otherwise the protection provided by the meter may be impaired 1 Before using the meter inspect the case Do not use meter if it is damaged the case or part of the case is removed Look for cracks or missing plastic Pay attention to the insulation around the connections 2 Inspect the test leads for damaged insulation or exposed metal Replace damaged test leads with identical model numb...

Страница 3: ...easurement Operation First check on 9V battery then turn rotary switch to the measuring position If the low battery sign will be displayed on LCD panel Nearly to sign on the meter front panel terminal input which also not exceed the testing voltage and current input value limitation 1 DC or AC Voltage Measurement See Diagram 2 1 Turn rotary switch to V or V voltage measurement 2 Insert the red tes...

Страница 4: ...Ω ohm measurement 2 Insert the red test lead into the Ω terminal and the black test lead into the COM terminal Connect the test leads across with the object being measured The measured value shows on the display The LCD displays OL indicating open circuit for the tested resistor or the resistor value is higher than the maximum range of the meter To maintain the resistance measurement accuracy disc...

Страница 5: ...The open circuit of diodes is 2 3V Do not input higher than DC 60V and AC 30V voltage to prevent any damage and accident 5 Temperature Measurements see diagram 7 1 Turn rotary switch to C 2 K type point contact temperature probe can only be used in the measurement below 230 C you need to separately buy another temperature probe 3 The LCD displays OL indicating point contact temperature probe which...

Страница 6: ...Current Range Resolution Accuracy 2000µA 1µA 1 Reading 2 Digits 20mA 0 01mA 1 Reading 2 Digits 200mA 0 1mA 1 2 Reading 2 Digits 10A 0 01A 2 Reading 5 Digits Overload Protection mA Range F2 fuse Ø6 35 x 31 8mm F 0 5A H 600V CE 10A Range F1 fuse Ø6 35 x 31 8mm F 10A H 600V CE Warning When 5A Continuous measurement is allowed When 5A Continuous measurement less than 10 seconds at an interval more tha...

Страница 7: ...Ω Buzzer beep continuously Overload Protection 600V AC or DC 7 Maintenance and Service Warning Make sure the test leads take off the terminal and the circuit as well as power off the meter if want to open the meter cover 1 General Service and Maintenance Periodically wipe the case with a damp cloth and mild detergent Do not use abrasives or solvents Any abnormal on the meter stop use the meter and...

Страница 8: ...selves the Information and the suitability of the products for their purpose and not make any assumptions based on information included or omitted Liability for loss or damage resulting from any reliance on the Information or use of it including liability resulting from negligence or where the Group was aware of the possibility of such loss or damage arising is excluded This will not operate to li...

Отзывы: