CALsys 250
20 |
P a g e
8.3 Monitoring the Controller Status
A row of beacons indicate the controllers status as follows,
OP1 Heat Output
OP2 Cool Output (only for models which operate below 0°C)
REM This beacon indicates activity on the PC interface
8.4 Units
Momentary pressing of the Scroll key will show the controller units °C or °F.
8.5 Setting the Input Type
a) Press ‘PAGE key’ for 4-5 sec. The display will show level 3.
b) Then press
▲
key, the display will show ‘CONF’. Then release
▲
key.
c) The display will ask for ‘CODE’. To enter the code for ‘CONF’ level, use
▼
&
▲
key. By default the CODE for ‘CONF’ is 4.
d) Then the display will show ‘INPUT’.
e) Then press scroll key (only once).
f) The 1
st
term, controller will display is ‘IN.TYP’. Here one can enter the type of
sensor we are using in respective Furnace like: RTD, Thermocouple (R, S, N, K
etc.). To enter the Input type use
▲
or
▼
key.
g) Press ‘PAGE’ and ‘scroll’ key together to return to ‘HOME’ display.
h) As we were in ‘CONF’, so change the level. Use
▼
key and select level 2.
i) After 2-3 second, the display will automatically show the ‘HOME’ display;
means temperature of furnace.
The upper display will show the current set sensor type,
J.tc
J thermocouple
K.tc K thermocouple
L.tc L thermocouple
Содержание CALsys 250
Страница 1: ...CALsys 250 1 P a g e Liquid Temperature Calibrator CALsys 250 USER MANUAL ...
Страница 10: ...CALsys 250 10 P a g e 3 2 Dimension ...
Страница 11: ...CALsys 250 11 P a g e 3 3 Wiring Diagram ...
Страница 25: ...CALsys 250 25 P a g e 9 2 Digital Communications Parameters The following table shows the parameters available ...
Страница 37: ...CALsys 250 37 P a g e NOTE ...
Страница 38: ...CALsys 250 38 P a g e ...