7. Operating Instruction:
7.1 Power
Plug the calibration bath power cord into mains outlet of the proper voltage, frequency
and current capability. Typically this will be (230 VAC±10, 50/60 Hz). Turn the
controller using the “Mains” switch located on the controlling section and set the
temperature value in the controller. After this, on the MCB which is provided in back
side of furnace. The black body will turn on and begin to heat the previously programmed
temperature set-point.
7.2 Heating up the source
Press “UP” or “DOWN” key of controller to change the set-point value. When the set-
point temperature is changed the controller will switch the calibrator heater on or off to
raise or lower the temperature. The displayed temperature will gradually change until it
reaches the set-point temperature. The Calibrator may require 20 to 30 minutes to reach
the set-point depending on the span. An another 10 to 15 minutes is required to stabilize
the bath temperature within ± 0.1 Deg C of the set-point.